3.2 Conductivity detector
■■■■■■■■■■■■■■■■■■■■■■
14
■■■■■■■■
2.945.0010 945 Professional Detector Vario – Conductometry
NOTICE
In order to prevent unnecessary peak widening after separation, the
connection between the outlet of the separation column and the inlet
to the detector should be kept as short as possible.
Connecting the detector inlet capillary to the separation
column
1 Connecting the detector inlet
■
Fasten the detector inlet capillary
sure screw (6.2744.070)
directly to the outlet of the column
1
2
3
Figure 6
Connection detector–separation column
1
Detector inlet capillary
2
PEEK pressure screw, short
(6.2744.070)
3
Separation column
Connecting the detector inlet capillary to the suppressor
1 Connecting the detector inlet
■
Connect the detector inlet capillary
suppressor
out
to each other using a coupling
and two short PEEK pressure screws