3.5 Connecting an Extension Module
■■■■■■■■■■■■■■■■■■■■■■
28
■■■■■■■■
942 Extension Module Vario ChS-PP (2.942.0300)
1
Prepare a tubing cartridge of the peristaltic pump for the regenera-
tion solution
2
Connect the capillary labeled
rinsing solution
to the outlet of the
pump tubing using a pressure screw (6.2744.070).
3
Connect the PTFE capillary from the rinsing solution bottle to the inlet
of the pump tubing.
3.5
Connecting an Extension Module
CAUTION
The 940 Professional IC Vario
has to be switched off
when connect-
ing the Extension Module!
Accessories
For this step you need the following accessories:
■
6.2156.060 cable Extension Module - Professional IC, 40 cm
or
■
6.2456.070 cable Extension Module - Professional IC, 1 m (optional
accessory)
The connection sockets are located on the rear of the instrument.
Connecting the Extension Module to the IC instrument
1
Plug the connecting cable (6.2156.060) into the
In
connection
socket on the Extension Module and tighten it in place.
2
Plug the other end of the connecting cable into the
Extension
Module
connection socket on the IC instrument and tighten it in
place.