3.8 Setting up accessories
■■■■■■■■■■■■■■■■■■■■■■
14
■■■■■■■■
NOTICE
If an incorrect sample vessel was inserted for the calibration with
the internal wavelength standard, the error
auto linearization
failed
may be displayed during the routine analysis.
2 Calibrating the instrument
Start the calibration in the control software
(see 8.105.8032 Tutorial
Vision Air Local and 8.105.8036 Tutorial Vision Air Network and
Server)
.
3 Repeating the calibration
Once a sample vessel with a different path length has been inserted,
repeat the internal calibration. To do this, start with Step 1 again.
For additional information on the calibration with the internal wavelength
standard, see
8.105.8032 Tutorial Vision Air Local or 8.105.8036 Tutorial
Vision Air Network and Server
.
3.8.2
Analyzing liquid samples
WARNING
Hot surface
Danger of burning due to hot components. The sample holder can be
heated to a maximum of 80 °C. After operation, the sample compart-
ment, the sample holder and the sample cup are very hot.
■
Wear protective gloves when handling sample holders and samples.
Fast and convenient analysis
Accessories
■
Suitable holder for disposable vials
■
Disposable vial
■
Lint-free cloth
■
Pipette
1 Preparing the vial
■
Add the sample liquid to the disposable vial using a pipette.
■
Wipe away sample residue and fingerprints from the outside of
the disposable vial using a lint-free cloth.