3.7 Temperature control
■■■■■■■■■■■■■■■■■■■■■■
12
■■■■■■■■
3.7
Temperature control
The DS2500 Liquid Analyzer supports a temperature range between 25 °C
and 80 °C. The minimum temperature that can be set must be 10 °C
above the ambient temperature T
min
= T
ambient
+ 10 °C
To achieve a temperature of 25 °C, the ambient temperature must be
15 °C. As the ambient temperature in most laboratories is approx. 25 °C,
Metrohm recommends to set the temperature to a minimum of 35 °C.
3.8
Setting up accessories
The following sample cups can be used with the DS2500 Liquid Analyzer:
Intended use
Holder for sample cups
Order number
Fast and convenient analysis
DS2500 holder for 2 mm disposable vials
DS2500 holder for 4 mm disposable vials
DS2500 holder for 8 mm disposable vials
DS2500 holder for 10 mm disposable vials
6.7492.000
6.7492.010
6.7492.020
6.7492.030
High precision
DS2500 holder for 1 mm cuvettes
DS2500 holder for 2 mm cuvettes
DS2500 holder for 5 mm cuvettes
DS2500 holder for 10 mm cuvettes
6.7492.100
6.7492.110
6.7492.120
6.7492.130
High sample throughput and automation
DS2500 holder for flow cells
6.7493.000
Calibration
DS2500 liquid wavelength calibration stan-
dard
6.7494.000
3.8.1
Calibrating the instrument
The DS2500 Liquid Analyzer is calibrated with an external wavelength
standard
(see "Calibration with an external wavelength standard", page
and with an internal wavelength standard
instrument with an internal wavelength standard", page 13)
WARNING
Hot surface
Danger of burning due to hot components. The sample holder can be
heated to a maximum of 80 °C. After operation, the sample compart-
ment, the sample holder and the sample vessel are very hot.
■
Wear protective gloves when handling sample holders and samples.