4.2 Setting up the titration cell for coulometry
■■■■■■■■■■■■■■■■■■■■■■
14
■■■■■■■■
899 Coulometer
4.2.3
Mounting the addition and aspiration tube (utilization with Ti
Stand)
6.1437.000
6.2713.000
6.2713.000
6.1439.010
1
2
4
5
6
3
Figure 6
Mounting the addition and aspiration tube
Insert the addition and aspiration tube as follows into the titration cell:
1
Attach the 6.2713.000 ground-joint sleeve that has been cut to size
to the ground joint of the 6.1437.000 stopper.
2
Insert the stopper into the 6.1439.010 addition and aspiration tube.
3
Attach the 6.2713.000 ground-joint sleeve that has been cut to size
to the ground joint of the addition and aspiration tube.
4
Insert this assembly into the ground-joint opening.
5
Connect the tubing for the reagent addition at the upper connector
of the addition and aspiration tube (5).