![Metrohm 2.872.0060 Скачать руководство пользователя страница 39](http://html1.mh-extra.com/html/metrohm/2-872-0060/2-872-0060_manual_1779508039.webp)
■■■■■■■■■■■■■■■■■■■■■■
5 Installation
872 Extension Module – Liquid handling
■■■■■■■■
31
3 Connecting the pressure side
NOTE
Depending on the use of the peristaltic pump, on the pressure side
you can either connect:
■
Case A:
a 6.2744.180 pump tubing connection
with filter
■
Case B:
a 6.2744.160 pump tubing connection
without filter
For pumping the auxiliary solutions to the MSM or to the SPM, a
6.2744.180 pump tubing connection
with
filter
must
be used.
Case A:
6.2744.180 pump tubing connection with filter:
1
2
3
Figure 20
Install pump tubing connection with filter
1
Union nut
2
Adapter
3
Tubing olive with filter holder
■
■
Select a suitable adapter
and slide it onto the pump tubing.
The type of adapter depends on the pump tubing
■
Place the tubing olive with filter holder
ing.
■
Screw the union nut
.
or
Case B:
6.2744.160 pump tubing connection without filter: