![Metrohm 2.872.0060 Скачать руководство пользователя страница 44](http://html1.mh-extra.com/html/metrohm/2-872-0060/2-872-0060_manual_1779508044.webp)
5.6 Injection valve
■■■■■■■■■■■■■■■■■■■■■■
36
■■■■■■■■
872 Extension Module – Liquid handling
3
Sample inlet
4
Sample outlet
5
Sample loop
Position A
In the position
FILL
, the sample solution flows
through the sample loop to the waste container.
Position B
In the position
INJECT
, the Dosino aspirates the
transfer solution through the sample loop to the
10-port valve. If sample solution is to be found in
the sample loop at the time of the valve switch-
ing, then this will be conveyed along with the
transfer solution, thus making its way via the 10-
port valve to the transfer tubing.