4.4 Eluent
■■■■■■■■■■■■■■■■■■■■■■
26
■■■■■■■■
872 Extension Module IC Pump
4 Installing tubing weighting and aspiration filter
1
2
3
4
5
Figure 16
Installing tubing weighting and aspiration filter
1
Eluent aspiration tubing 6.1834.080
2
Eluent bottle attachment 6.1602.160
3
Tubing weighting
From accessory set 6.2744.210.
4
Clamping screw
From accessory set 6.2744.210.
5
Aspiration filter 6.2821.090
With filter holder from accessory set
6.2744.210.
■
onto the eluent aspiration tub-
ing
.
■
Slide the clamping screw
onto the eluent aspiration tubing
.
■
Insert eluent aspiration tubing
into the aspiration filter
. The end of the tubing must touch the base of the filter.
■
Screw together clamping screw
The end of the tubing must still touch the base of the filter.
Figure 17
Eluent aspiration tubing fully equipped.
5 Mounting eluent aspiration tubing to the eluent bottle
■