![Metrohm 2.870.0010 Скачать руководство пользователя страница 38](http://html1.mh-extra.com/html/metrohm/2-870-0010/2-870-0010_installation-and-tutorial_1779535038.webp)
3.2 Carrying out titrations
■■■■■■■■■■■■■■■■■■■■■■
32
■■■■■■■■
870 KF Titrino plus
■
Select the menu item
Calculation
and open the corresponding
dialog with
[OK]
.
The calculation formula is defined for every method template and
cannot be modified. Depending on the unit of the sample size
and the result, only the conversion factors
FCT
and
DIV
need to
be adjusted.
■
Select the line with the calculation formula and press
[OK]
.
This conversion table is for purposes of information only. It is not
possible to apply values taken from there.
■
Find the desired result unit and the unit of the sample size in the
table and make a note to yourself of the factor and divisor con-
tained in this line.
■
Close the table with
[BACK]
.
■
Enter the values for
Factor (FCT)
and
Divisor (DIV)
according to
the conversion table shown above.
4 Activate the statistics
■
With
[BACK]
, switch to the parameters dialog.
■
Select the menu item
Statistics
and open the corresponding dia-
log with
[OK]
.
■
Select the parameter
Statistics
and confirm with
[OK]
.
■
Select
on
and confirm with
[OK]
.
The parameter
Number of determinations
is additionally dis-
played.
■
Select the parameter
Number of determinations
and open the
editing dialog with
[OK]
.
■
Enter the number of determinations to be carried out, in this
example
3
.
■
Apply the value with
Accept
or
[BACK]
.