![Metrohm 2.850.9010 Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/metrohm/2-850-9010/2-850-9010_manual_1779682020.webp)
2.4 Connecting the detector inlet capillary
■■■■■■■■■■■■■■■■■■■■■■
12
■■■■■■■■
2.850.9010 IC Conductivity Detector
Connecting detector inlet capillary to MSM / MSM
‑
HC
1 Connecting detector input
■
Connect the detector inlet capillary
and the capillary of the
MSM / MSM
‑
labeled with
out
to one another using a
6.2744.040 coupling
and two 6.2744.070 short pressure
screws
1
2
3
4
Figure 5
Connector detector – MSM
1
Detector inlet capillary
2
MSM / MSM
‑
HC outlet capillary
labeled with
out
.
3
Coupling 6.2744.040
4
Pressure screws, short 6.2744.070
Connecting the detector inlet capillary to the MCS
1 Connecting detector input
■
Fasten the detector inlet capillary
PEEK pressure screw
.
3
2
1
Figure 6
Connection detector – MCS
1
Detector inlet capillary
2
Pressure screw, long 6.2744.090
3
MCS output
Содержание 2.850.9010
Страница 1: ...IC Professional Detector IC Conductivity Detector Manual 8 850 8057EN 2019 09 23...
Страница 2: ......
Страница 6: ...Table of contents IV 2 850 9010 IC Conductivity Detector 7 Accessories 19 Index 20...
Страница 8: ......