M1 – M3  Display units 

 

Page 92 

 

 

 

 

 

 

15.  FACTORY RESET 

 
This function allows to come back to the factory setting. 

 
Warning: At the end of this procedure, all the parameters of your M3 will be reset 
(tolerances, masters, formulas…)  

 
To Reset the M3 to its factory settings, please follow the process bellow. 
 
1 – Shut down the M3 
2 – Start the M3 
3 – When the splash screen appears, press on the Metro logo 
4 – A desktop with 3 icons appear.  
  

4a – Press on « initialization » 

  

4b – Confirm YES or cancel NO  

5 – Press Home to come back to the measuring screen 
 

 

 

 

 

 

 

4a 

4b 

Содержание M1

Страница 1: ...y u un ni it ts s f fo or r 1 1 o or r 2 2 a ai ir r g ga ag ge es s USER S MANUAL Firmware 3 00 33 0 450 39 08 49 Metro Fax 33 0 450 39 08 33 80 impasse des chapotines web www metro fr com ZAE Chez Merlin F 74420 St André de Boëge France E mail info metro fr com ...

Страница 2: ...HE USB STICK CONNECTOR 21 3 3 12 THE FOOTSWITCH CONNECTOR 21 4 AIR PREPARATION UNIT AIR PIPES _____________________________ 23 4 1 AIR PREPARATION UNIT 23 4 2 AIR PIPES 23 5 QUICK START _________________________________________________ 25 6 GRAPHICAL INTERFACE ________________________________________ 26 6 1 2 MAIN PARTS 26 6 2 GENERALITIES 27 6 3 CONFIGURATION WINDOWS 29 6 4 VIRTUAL KEYBOARD 30 7 ...

Страница 3: ...SURING SCREEN 72 11 DATA EXPORT ON USB KEYS____________________________________ 73 11 1 STANDARD MODE 73 11 2 ADVANCED MODE 74 12 CONFIGURATION OF THE UNIT WITH QR CODES ___________________ 75 12 1 CALIBRATION COMMANDS 75 12 2 ADVANCED CONFIGURATION 76 12 3 EXAMPLE OF QR CODE CONFIGURATION 76 13 MODBUS RTU Protocol __________________________________________ 78 14 OPTIONNAL I O MODULE______________...

Страница 4: ...ge caused by the use of non supplied products components or accessories It also does not cover products modified without any written approval from the manufacturer including electrical or mechanical modification removal of serial numbers or of the manufacturer s trademarks or of any other identification THE SOLE RECOURSE UNDER THIS GUARANTEE SHALL BE THE REPAIR OR THE REPLACEMENT OF DEFECTIVE PART...

Страница 5: ...erformances and the use of this product It is the responsibility of the user to verify the calibration of the display before measuring and it is advised to check periodically the calibration and measurement performance CLEANING Use a soft cotton cloth slightly soaked with an ethyl alcohol based product DO NOT USE the following products acetone benzene toluene and halogens hydrocarbons ...

Страница 6: ... characteristics on the screen and can make static measurement or dynamic Max Min Max Min The M1 M3 can be connected to a PC thanks to its RS232 or USB connection A footswitch can be connected in order to transfer measurements The M1 M3 displays for air gages requires 2 eventually 3 masters for calibration 3 2 VERSIONS REF Description 11040 Display unit M1 for 1 air gage 13040 Display unit M3 for ...

Страница 7: ...ation with a PC USB port for communication and or power supply USB Stick for data saving on a CSV file Optional connection of M Bus modules Measurement transfer by pressing a key footswitch input or retro command on the RS232 port Operating temperature 15 C to 30 C Power supply from 85 to 265 VAC by using the supplied main transformer or by connecting it directly on your PC USB port or through the...

Страница 8: ...also possible to panel mount the display with the accessory ref ACS AFF 001 Be careful not to use longer screw than M4 16 Pneumatic connections see also chap 4 Remove the antiskid parts to attach it on the table or on a vertical panel ...

Страница 9: ...M1 M3 Display units Page 9 ...

Страница 10: ... includes 1 M1 3 display mounted on an orientable stand 1 USB cable length 1 8m for power supply and or data transfer 1 USB main adaptor for the display power supply The M3 can also be powered by a computer when connected to it with the USB 1 Mini CD containing the user manual ...

Страница 11: ...NE 008 For M3 displays It allows to divide the 8mm tube at the output of the air preparation unit into 2 tubes of 6 mm adapted for the connection on the M3 input ACS PNE 006 Soft polyurethane tubes 6 4mm The M1 M3 are fitted with 2 connectors for air tube with 6mm external diameter and 4mm internal diameter We advise to use soft polyurethane tubes from SMC because their flexibility prevent the dis...

Страница 12: ... 8 input outputs isolated with optocouplers allowing to get additional functionalities for example output for Go noGo input for preset or start dynamic measurement The M Bus modules are mounted on an aluminum profile allowing to mount them on a standard DIN rail Up to 4 modules can be mounted MB RL The optional MB RL module is fitted with 2 independent relays min and max free of potential that ind...

Страница 13: ...o connect the compatible M Bus modules on a M3 display unit Length 1 5m 45160 RS232 cable This cable allows to connect a M1 M3 display to a computer or a PLC 18193 This cable allows to connect a M1 M3 display to a Multiplexer Mux from Metro and to the DP1 printer from Mitutoyo ...

Страница 14: ... value of the M1 M3 Requires cable ref 45179 to be connected to the M1 M3 ACS AFF 003 QR code reader for Metro displays 2 main functions Allow to send a configuration on the display part definition tolerance display type calibration of the air gage etc Send a Manufacturing order number The export file will take the name of the number entered ACS PNE 012 Air cut off footswitch Allows to simply cut ...

Страница 15: ...OVERPASS 4BAR 0 4MPA It is very recommended to use the FRL Air preparation set from Metro ref ACS PNE 003 See details on the chap 3 3 4 RS232 connector for connection to a PC Multi function footswitch connector Mini USB connector for power supply and measurement transfer to a PC To connect m bus modules optional 24 VDC power supply If you choose this connector to power the display the ON OFF switc...

Страница 16: ...restrictor see chap 3 3 8 But if you want to use the M1 M3 display together with your existing air gage you will have to use the adapted restrictor by your own It is advised to contact Metro or one of its distributors for advices or confirmation around this subject Check this video for understanding the influence of the restrictor on the linearity link to the video Or scan the QR code Nozzle diame...

Страница 17: ...M1 M3 Display units Page 17 ...

Страница 18: ...his configuration allows to have a faster the reaction time and makes the installation easier by removing the need to choose the adapted restrictor If you order a turn key solution at Metro it will generally be delivered according to this principle The cabling schema is the following Air gage with by pass nozzle integrated restrictor Regulation unit ref ACS PNE 003 delivers 3BAR 0 3MPa pressure St...

Страница 19: ...guration is as following 9600 bauds 8 bits 1 stop bit no parity CONNECTOR PINOUT It is fitted with a SUBD 9 pins female connector Pin Signal Direction Description 1 Not used 2 RX Input Reception of data 3 TX Output Transfer of data 4 IN1 Input Do not use Only for firmware update 5 Gnd Ground 6 Not used 7 IN2 Input Do not use Only for firmware update 8 9 Not used ...

Страница 20: ...d with the standard drivers of your operating system Windows Mac OS etc When you will send the measurement the value will be written on your PC screen where your cursor is in the same way as it would have been typed with a keyboard Message that appears when the M1 M3 has been correctly installed 3 3 11 The 24VDC connector It is advised to use this power supply when the M3 is panel mounted Using th...

Страница 21: ...ded on the CSV file a output csv file is created when the operator transfers the data for the first time When the USB stick has been correctly detected a USB logo will appear on top bar If the USB stick is not connected when the operator transfers the measurement an error message will pop up USB sticks with partitions are not supported 3 3 12 The footswitch connector This connector is used to conn...

Страница 22: ...he displayed part reference Hold freezes the display In this case the button CLEAR in the measuring screen becomes yellow Screenshot require to have a footswitch connected a USB stick inserted It creates a BMP file screenshot in the USB stick ...

Страница 23: ... display consist of 1 push in connector for 8mm ext tubes 1 shut off valve 1 regulator adjusted at 3 BAR 0 3 MPa with a 1µm filter this pressure is absolute 1 precision regulator adjusted at 3 BAR 0 3MPa 1 push in connector for 8mm ext tubes 4 2 Air pipes The M1 M3 are fitted with air connectors adapted to connected pipes with 6mm external diameter and 4mm internal diameter In the case of the M3 i...

Страница 24: ...M1 M3 Display units Page 24 ...

Страница 25: ...Repeat the step B for the MAX master D You can check by placing again the MIN master The value on the position field must be at 0 0000 then by placing the MAX master the position field must indicate the dimensional difference between the 2 masters 6 Configuration of the part to be measured From the Icon definition A Use the scroll bar to reach the 2nd screen B Write the value of MIN and MAX tolera...

Страница 26: ...parts 1 A part that allows configuring the device and the measure It consists of an icon desktop with windows Icon desktop Icon desktop with configuration windows The second part measuring screen can be reached by pressing the button 2 This part allows to see the measurement result and to use them The M3 starts on this screen For reaching the configuration screens press on the button Example of me...

Страница 27: ... to enter a characteristic name either with an RS232 command or by scanning a QR code After this the characteristic name is written instead of P1 C 1 Name of the active part reference an formula Here Part 1 with Channel 1 Static or dynamic Max Min etc Unit mm µm or inch Inside or outside measurement Characteristic name entered with RS232 or QR code ...

Страница 28: ...ing a QR code Note When either a characteristic name or a manufacturing order name has been entered only 1 information appear on the blue bar However the information display can be changed by touching the area Manufacturing order number entered with RS232 or QR code USB logo appear when a USB stick or a QR code reader has been plugged ...

Страница 29: ...and are saved after validating while quitting the window Here after are the different ways to input data Multiple selection box Press on the black arrows to change the pre defined value Edit box A virtual keyboard appears after clicking on the edit box Closing a window All the windows can be closed by clicking on the white cross on a red background on the top left corner on each window Closing a w...

Страница 30: ...M1 M3 Display units Page 30 6 4 VIRTUAL KEYBOARD Erase Quit without taking into account the modification Quit and save the modification ...

Страница 31: ...e communication etc from this window The measure tolerances characteristics etc is also configured from this window The 6 following sections describe the 6 icons of this screen Configuration of the device language footswitch function etc Allows to lock selected functions by password Go to the measuring screen Display mode selection 1 or 2 bargraph needle without tolerance Characteristic edition to...

Страница 32: ... gives the possibility to define the tolerances the master the master part characteristic the resolution of each of the 2 available measurement configurations This window is divided in 2 parts for the configuration of the 2 measurement configurations Go to the screen 1 2 3 or 4 Choice the part 1 or 2 ...

Страница 33: ...an For measuring internal or external diameters It defines the indicators colour and gives information to the user if the part can be used again Choice of the calculation formula C 1 C 2 C 1 C 2 C 1 C 2 C 1 C 2 C 1 C 2 Taper C means Channel It is possible to give a coefficient for each probe from the menu ...

Страница 34: ...automatic choice when the formula is set as Taper DMS Degrees Minutes Seconds Mode Static direct reading The display is refreshed together with the probe is moving Dynamic The following mode starts after pressing on the clear button of the measuring screen or through a I O or footswitch command Max Min Max Min Average Median max min 2 ...

Страница 35: ...mits will appears in red All the good measure will be in green To measure the angle you require an air gage with 2 level The distance between each diameter has to be entered on the parameters When the taper mode is used the choice of the unit is limited to deg decimal degrees or DMS degrees minutes seconds A new choice appears nozzle The distance between the 2 diameters must be written here ...

Страница 36: ...OW In this case the internal diameter is used as on the pictures bellow On the measuring screen a yellow indicator shows when the part is too small so possible to rework and a red indicator if the part is too big rejected REWORK PASS REJECT Direction External diameter The opposite phenomena happens for the external diameter REJECT PASS REWORK 7 1 2 Part 2 This screen allows to input tolerances and...

Страница 37: ...allows the user to sort the parts according to their position in the tolerance interval The top bar of the display which is normally blue is grey on the picture bellow because grey is the color which was defined for this interval of measurement Choice of the number of class Choice of the color for each class The limit defined for each class are selected by default but they can be changed by the us...

Страница 38: ...M1 M3 Display units Page 38 Grey color because of the class parameters ...

Страница 39: ...its warnings and to define them Control limits are warning that informs the user is the measure becomes too close to the tolerances limits Yellow colour on the bargraph This part is hidden if no has been selected Example of characteristic displayed with control limits Part OK Part OK with warning Part Not OK ...

Страница 40: ...edle galva horiziontal bargraph or value only The auto switch function allows to call the corresponding program by a simple probe movement Example Move probe 1 part 1 is displayed Move probe 2 part 2 is displayed Left simple and double Galva cursor None simple and double The keys appears on the needle indicator galva and allow to move the needle like with the screw on an old analogue display ...

Страница 41: ...ring screen only the bargraph appears Displays small indicators triangles on the measuring screen to show the maximum and minimum value measured Allows to move the needle like with the screw on an old analogue display Allows to display 2 probe on the measuring screen except in GALVA mode ...

Страница 42: ...ring position and the dimensional difference between the 2 masters when the MAX master is inserted Channel 1 and channel 2 only for M3 Possibility to connect an M Bus Allows to slow down the refresh frequency of the display running average It can be useful to limit the blinking of the last digit and for runout measurement on large parts If you touch the unit it can switch to BAR and MM only availa...

Страница 43: ...on of the air gage 7 3 1 1 Calibration in 2 points The air gage probes cannot be used without a special calibration This calibration requires at least 2 master sample of the size of the tolerance limit for the part to measure Must be calibrated ...

Страница 44: ...M1 M3 Display units Page 44 7 98 8 02 Write here the 2 calibration values which will be used ...

Страница 45: ...M1 M3 Display units Page 45 After placing the min master push calibration The same manipulation is done here but with the max master and the final push on calibration The probe is now calibrated ...

Страница 46: ...imple sorting Go NoGo the results will be correct but if you need to have the real value and or to use the class sorting functionality you need to use the most adapted restrictor see chap 3 3 1 7 3 1 2 Calibration in 3 points There is the possibility to use 3 points to calibrate the measure When the button central is selected a third master point is asked The calibration process stays the same wit...

Страница 47: ...M1 M3 Display units Page 47 In measuring mode if the input pressure has changed over the limit a popup windows appear informing the operator and asking the recalibrate or check the air network ...

Страница 48: ...en It allows to mix 2 different probes on the same display with the multiplexer modules or to use input output modules Multiplexer modules The same process is made to connect all the multiplexer modules Example M3 with MB 2S module for Heidenhain probes on the channel 2 The probe associated with each channel can be chosen here ...

Страница 49: ... module must be configured from the C1 or C2 pages of the setup menu according to the channel with which the probe is linked Please refer to the example for each probe chapter 5 3 to 5 3 6 for air gage MB 2S module Heidenhain MB 2T module TTL MB AG Module Air Gage ...

Страница 50: ...M1 M3 Display units Page 50 MB 8I Module Inductif MB 4D Module For Digimatic insturments MB 1R Module RS232 ...

Страница 51: ...ction of an input output module is made with the same way than a multiplexer Up to 4 MB IO can be connected Example connection of a MB IO module on the M3 The different options for the MB IO configuration are displayed bellow Click here to configure the MB IO ...

Страница 52: ...ed value to 0 relative mode The number of the MB IO is written here and can be changed if more than one is connected These buttons are used to output a 24VDC signal on specific port of the MB IO to check the installation At the opposite if you input a 24VDC on a terminal the corresponding button becomes green The function of each port of the MB IO can be set with this 2 menus The first one is for ...

Страница 53: ... out of the lower tolerance limit No Go output output a signal if the part is out of the upper tolerance limit Part status output output a signal if the whole part the 2 probes respects the tolerance limit Class output output a signal if the part is in the selected class Hold Input freeze the screen to display the previous measure as long as the input signal stay Characteristic output changes the ...

Страница 54: ... RS232 Keyboard or USB Stick Footswitch function Preset Calibration transfer send the measurement to the RS232 or USB part change the active part part 1 part 2 Init Dyn start dynamic measurement min max zero Set the displayed value at zero hold freeze the measuring screen the clear button becomes yellow when the screen is frozen screenshot generate a BMP file on the USB stick when the footswitch i...

Страница 55: ...ou select transfer USB Stick a output csv file will be created on the connected USB stick and each time you press the Print button of the measuring screen a line will be added on the file If you use a QR code reader to input a manufacturing order number the OUTPUT CSV file takes the name becomes manufacturingordernumber csv The clear icon is yellow while the hold function is activated clear button...

Страница 56: ...M1 M3 Display units Page 56 Defines the register format for the MODBUS RTU protocol Float or integer Ask a confirmation of preset when the preset is done from the measuring screen ...

Страница 57: ...tions of the M3 If locked a password will be asked to access to the icon desktop Modification of the password By default 0000 Allows to remove the corresponding buttons from the measuring screen It is recommended to remove the button to make the interface as easy as possible for the user Example of a measuring screen without all the buttons ...

Страница 58: ...M1 M3 Display units Page 58 Allows to remove the corresponding buttons from the measuring screen Allows to lock the measuring mode maxi mini maxi mini average ...

Страница 59: ...M1 M3 Display units Page 59 7 6 MEASURE After clicking on this button the measuring screen appears Please read the chapter 8 for the presentation of the measuring screen ...

Страница 60: ...comes green for one second Absolute or relative mode allows to make a zéro at a defined position If the relative mode is selected the button become red To come back to the absolute mode press on the clear button Calibration Preset After pressing this button the displayed value will take the master value Measurement transfert RS232 or USB or USB stick Change manually the active part part 1 or part ...

Страница 61: ...tton is used to set the current measured value to 0 it s the relative mode Clear This button allows to cancel the zero button and to switch again to absolute mode Preset This button allows to set the current measure value to the master value ...

Страница 62: ...M1 M3 Display units Page 62 Print The print function allows to transfer the measured value to a computer Part This function is used to change the displayed part ...

Страница 63: ...lva see chap 7 2 it is possible to change manually the type of scale Automatic Manual The automatic needle scale adapts automatically with the part tolerances The manual needle has fixed scales and allows to use the option see chap 5 2 You can change the scale type by touching the area as on the picture bellow during 2 seconds ...

Страница 64: ...ssible to change temporarily to a dynamic mode directly from the measuring screen To change the measuring mode press on the icon as shown on the above picture Static Max Min Max Min Average only adapted on an automatic fixture Median The displayed value is calculated with the formula MAX MIN 2 When a dynamic mode has been selected press on the button or on the footswitch if configured as init dyn ...

Страница 65: ... this screen Press on the value as shown on the below picture an input the new value with the keyboard If the double display mode is selected the 2 characteristics can move together but only 1 is active The lateral buttons have an effect only on the active characteristic For selecting the active characteristic press on it and the top of the box will become green The lateral button have an effect o...

Страница 66: ...n you transfer the measure with screen through IO or footswitch the displayed value will appear on your PC when your cursor is For example on an Excel cell in the same way than you would have typed with your standard keyboard The communication with an USB cable is made by selectioning KeyBoard communuication type The M3 will communicate with the computer as would a key board do Example of a table ...

Страница 67: ...board so in order to write numbers the num lock key must be activated If your keyboard doesn t have the num lock key you can still find it by writing visual keyboard in the windows search bar and activate it If even there the num lock key doesn t appear the option key allows to access to a menu from which you can activate the num lock ...

Страница 68: ...ctions For connecting the M3 to a PC or a PLC you must use a cable Metro ref 45160 sold separately 10 1 Commands 10 1 1 Generalities All the commands must be ended by a CR character ASCII code 13 0D The commands can be sent grouped by separating with a max 500 characters The display is refreshed only once after reception of the CR character ...

Страница 69: ...atic x 1 maxi x 2 mini x 3 maxi mini x 4 average x 5 median Formula nFM nFM x x 0 C1 x 1 C2 x 2 C1 x 3 C2 x 4 C1 C2 x 5 C1 C2 x 6 C1 C2 x 7 C1 C2 Direction nDIR nDIR x x 0 sans x 1 internal x 2 external Upper tolerance nUT nUT seee ddddd Nominal nNM nNM seee ddddd Lower tolerance nLT nLT seee ddddd Master nMT nMT seee ddddd Activate control limits nLIMIT nLIMIT x x 0 inactive x 1 active Upper cont...

Страница 70: ...only no tolerance Display DISPL DISPL x x 1 display 1 characteristic on the screen x 2 display 2 characteristics on the screen 10 1 2 2 Window CONFIGURATION Window Configuration Rubric Read command Write command Comment Transfert PRINT PRINT x x 0 USB x 1 RS232 Langue LANG LANG x x 0 French x 1 English x 2 German x 3 Spain x 4 Italian x 5 Hungarian x 6 Czech x 7 Swedish x 8 Portugaise Footswitch F...

Страница 71: ...71 x 3 change Part reference x 4 start dynamic measurement 10 1 2 3 Window Lock Window Lock Rubric Read command Write command Comment Protection LOCK LOCK Lock UNLOCK Unlock Code PASS PASS xxxxxx xxxxxx password with 6 figures ...

Страница 72: ...he value with format 000 00000 000 00000 CR 1 Transfert the characteristic 1 format 000 00000 CR 2 Transfert the characteristic 2 format 000 00000 CR Part G1 Part 1 G2 Part 2 M3 power off OFF Shut down the M3 Reset the M3 RST Reset the unit remove the configuration calibration etc Factory reset FCTRST Deep reset Gives a manufacturing order number OFA xxxxxxxxxxxx The manufacturing order number app...

Страница 73: ...configure the menu CONFIGURATION COMMUNICATION USB Key Then when the transfert is triggered by footswitch touch of the PRINT button by an input action or Modbus a file will be created on the USB key 11 1 Standard mode As a standard a file named output csv is created on the USB key It look likes this Output csv Characteristic 1 2 COMMENT Name P1 C 1 P2 C 2 Name of the characteristics PART x CHANNEL...

Страница 74: ...een and the output csv file will get the name of the scanned name Then each time you scan a new number after saving the first part a file is created on the USB key The command to encode the QR code is OFA xxxxxxxxxxxx Example Standard screen before scanning manufacturing number After scanning the following QR code corresponding to the command OFA ORDER 12345 The screens become like this Then when ...

Страница 75: ...ing to the following layout Instructions Corresponding QR Code AGI Initialisation AGL 1 Configuration of input 1 AGF 1 Filter level from 0 to 5 AGN 20 00000 Value of the MIN master AGC 20 01000 value of the CENTRAL master optional AGA 20 02090 Value of the MAX master AGS End of calibration configuration display of the configuration window If you want to configure 2 channels same principle but you ...

Страница 76: ...me This command is RST After sending this command you must start the configuration by the general parameters then configuration of the characteristics and at the end sending the calibration commands as described on the previous chapter 12 3 Example of QR code configuration The following example shows a complete configuration of the unit with QR code We use 2 air gages 1 plug and 1 ring mounted on ...

Страница 77: ...master AGL 2 Configuration of input 2 AGF 0 Filter level from 0 to 5 AGN 19 9807 Value of the MIN master AGA 19 9932 Value of the MAX master AGS End of calibration configuration display of the configuration window After this configuration a popup window appears asking DEACTIVATE CALIBRATION YES NO Press yes and the previous calibration will be erased then the window for calibrating the display ope...

Страница 78: ... codes 3 by reading and 16 by written Data 1 to 256 register CRC16 verification code on 2 octets The following functionality are available Reading of the position of 2 probes Instant reading of the value of 2 characteristic Preset Reading programing of the tolerances of the master value the formula Register are compounded by one or a couple of 16 bits words Function Address Size word Value Preset ...

Страница 79: ...1 1 à 16 Unit 110 1 0 mm 1 inch 2 µm Control limit activated 111 1 0 no 1 yes resolution 112 1 1 to 5 decimal State of the characteristic 113 1 0 good 1 Mini 2 Maxi 3 lower ctrl limit 4 upper ctrl limit Type of the characteristic 114 1 0 Static 1 Maxi 2 Mini 3 Maxi Mini 4 average 5 Median Nominal 115 2 Lower tolerance 117 2 Upper tolerance 119 2 Master value 121 2 Measure 123 2 Displayed value Min...

Страница 80: ...56 2 Maxi class 1 158 2 Maxi class 2 160 2 Maxi class 3 162 2 Maxi class 4 164 2 Maxi class 5 166 2 Maxi class 6 168 2 Maxi class 7 170 2 Maxi class 8 172 2 Maxi class 9 174 2 Maxi class 10 176 2 Maxi class 11 178 2 Maxi class 12 180 2 Maxi class 13 182 2 Maxi class 14 184 2 Maxi class 15 186 2 Maxi class 16 188 2 probe 1 7000 2 probe 2 7002 2 ...

Страница 81: ...4 M Bus modules with 8 inputs output isolated by opto coupler ref MB IO The 8 outputs are similar to the open collector PNP type They can be used with an external power supply 12 to 30 VDC maximum The maximal output current drained by each output is 50mA The 8 inputs represent a 2 2kOhms load connected to the 0 volt ...

Страница 82: ...r control limit Output a signal if the part is in the upper control limit Out Tol Output a signal if the part is out lower tolerance Out Tol Output a signal if the part is out upper tolerance Part status Output a signal if the part is in the tolerances Class Output a signal on a defined port for each class Stop Hold the measure characteristic Change the displayed characteristic Touchscreen Disable...

Страница 83: ...s Page 83 Connection example between a M3 and a PLC Output to the PLC 0V of the PLC E1 0 V IO module PLC 24V Output 24 V of the PLC 0 V IO module PLC 0 V 0 V 0V of the PLC PLC Output 24V E1 Input 24 V from te PLC 24V 24V ...

Страница 84: ...a M Bus module The MB RL module has been designed for retrofitting old Monocote displays equipped with the optional relay board The external M bus module has the same pinout SUBD 15 than the relay board of the Monocote display so there is no need to change the wiring of the existing installation ...

Страница 85: ...mories for the dynamic measurement It must be used every time the dynamic measurement starts when the part to be measured is already in place under the probes This command has the same function as the button CLEAR of the measuring screen STOP This input stops the measurement and freezes the screen as long as it is maintained at the logical level 1 ZERO This input controls the zero reset of the dis...

Страница 86: ...ibration command 11 MAXNC Maxi tolerance relay contact 1A 48V 12 MAXCOM Common of maxi tolerance contacts 13 MAXNO Maxi tolerance relay contact 1A 48V 14 INIDYN input Command for measurement initialisation 15 ZERO input Command for display reset to zero 12V STOP 2 2k Masse E gnd Monocote Commande STOP 2 2k E gnd Monocote Commande 24 V 0 V Command example with internal supply not isolated and with ...

Страница 87: ...hanged Input start a dynamic measure Output output a signal if the part is out of the tolerance max Input freeze the measurement screen and the displayed value Input Set the displayed value to zero Input Transfer the measured value need a USB or RS232 link Input Set the displayed value to the master value Output output a signal if the part is out of the tolerance min ...

Страница 88: ...ions are then available alarm or compensation A Identification and configuration of the Temperature module MB TP B Configuration of the temperature mode After the identification described on the previous chapter a new window is available from the last page of the menu DEFINITION The temperature is displayed Selection between type K or PT100 Unit Degree Celsius or Fahrenheit ...

Страница 89: ...C of allowed temperature variation Let s come back to the measure screen After a PRESET the actual temperature is recorded and the monitoring starts Temperaure OK no alarm Temperaure increased alarm After validation with the OK button the Preset button becomes red until you made the preset The PRESET button become black again after preseting The new reference temperature has now been moved to the ...

Страница 90: ...d 14 Nickel 13 Iron or Steel 11 1 Carbon steel 10 8 Platinium 9 Glass 8 5 Tungsten 4 5 The compensation is done assuming that the Master part has been measured at 20 C If it is not the case select the option advanced and a next page appear allowing to enter the temperature of the master during the measurement according to its certificate Compensation mode Expansion coefficient If validated green a...

Страница 91: ...advanced mode you need to enter the temperature of the certificate and the reference temperature For example if you enter a certificate temperature of 20 and a reference temperature of 30 when you PRESET the displayed value will represent the size of the master at 30 The temperature function is advanced functionality of M3 It is recommended to make trials and comparative measurements to validate t...

Страница 92: ...f your M3 will be reset tolerances masters formulas To Reset the M3 to its factory settings please follow the process bellow 1 Shut down the M3 2 Start the M3 3 When the splash screen appears press on the Metro logo 4 A desktop with 3 icons appear 4a Press on initialization 4b Confirm YES or cancel NO 5 Press Home to come back to the measuring screen 3 4a 4b ...

Страница 93: ...o not have a Metro cable ref 18060 you can make your own cable by following this schematic The firmware update requires to use the flash magic software that can be download free from this address www flashmagictools com After installation please configure the software on the following way M3 Sub D9 male 1 Rx 2 Red Tx 3 Blue DTR 4 Green GND 5 Yellow 6 RTS 7 White 8 9 PC Sub D 9 female 1 CD 2 Rx 3 T...

Страница 94: ...m 2 to 5 min depending on the Baud rate selected During the update the screen becomes blurred 5 The M3 restart automatically when it is finished 6 Reset the M3 by following the procedure on the chapter 11 The LPC4357 can be found on the ARM Cortex category after clicking on Select If you are connected directly on a serial port you can use the baud rate 115200bauds If you use a USB RS232 convertor ...

Отзывы: