Entry of Equipment & Conditions
Metrix recommends that the manufacturer, model, serial number, calibration date and calibration due date
of all test equipment be documented on the test report. In cells C40 through F44 users can enter this data
for the accelerometer, signal conditioner and volt meter.
It is also good practice to note the name of the technician, date and laboratory conditions if possible. This
data can be entered into cells H7 through H10.
7