Power and Voltage Quality Trainer
AD 2
– MI 3399
165
Figure 4-34: General recorder setup
The instrument is now prepared for voltage event recording, which is generated by the
Application Trainer. Please start the recorder by pressing the
F1 (“START”) key. The
recorder can be stopped anytime, by pressing the F1 (STOP) key.
4.4.2 Observing interrupts on the instrument
Enter the RECORDERS
EVENTS TABLE menu and check recorded events. The
events table should be similar as shown in the figure below.
Figure 4-35: Event table (interrupts)
Additionally, the waveform shape of a recorded event can be observed from the
RECORDERS
MEMORY LIST, by selecting an appropriate recorded waveform and
observing it, by pressing the F1 (VIEW) key. The recorded waveform should be similar
to the one shown in the figure below.
Figure 4-36: Voltage interrupt RMS measurements and waveform shape
Содержание MI 3399
Страница 27: ...Electrical Installation Safety Trainer AD 1 MI 3399 27...
Страница 28: ...Electrical Installation Safety Trainer AD 1 MI 3399 28 AD 1 MI 3399 EIS Electrical Installation Safety Trainer...
Страница 46: ...Electrical Installation Safety Trainer AD 1 MI 3399 46 3 Exit settings...
Страница 74: ...Electrical Installation Safety Trainer AD 1 MI 3399 74...
Страница 122: ...Power and Voltage Quality Trainer AD 2 MI 3399 122 AD 2 MI 3399 PQA Power and Voltage Quality Trainer...
Страница 197: ...Power and Voltage Quality Trainer AD 2 MI 3399 197 Figure 4 89 Transient Chart view...
Страница 204: ...PAT Appliances Machines Safety Trainer AD 3 MI 3399 204...
Страница 205: ...PAT Appliances Machines Safety Trainer AD 3 MI 3399 205 AD 3 MI 3399 PAT Appliances Machines Safety Trainer...
Страница 249: ...PV Photovoltaic Systems Trainer AD 4 MI 3399 249...
Страница 250: ...PV Photovoltaic Systems Trainer AD 4 MI 3399 250 AD 4 MI 3399 PV Photovoltaic Systems Trainer...