MI 2792 PowerQ4 Plus
Operating the instrument
72
3.16.1 Connection setup
Figure 3.54: “CONNECTION SETUP” screen
Table 3.78: Description of Connection setup
Nominal range
Nominal voltage range. Select voltage range
according to the
nominal network
voltage.
1W and 4W
3W
50 ÷ 110V (L-N)
86÷190 V (L-L)
110 ÷ 240V (L-N)
190÷415 V (L-L)
240 ÷ 1000 V (L-N)
415÷1730 V (L-L)
Note:
Instrument can accurate measure at least 50%
higher than selected nominal voltage.
Voltage ratio
Scaling factor for voltage transducer.
Use this factor if external voltage transformers or
dividers should be taken into account. All readings
are then related to the primary voltage. See 4.2.2 for
connection details.
Note:
scale factor can be set only when the lowest
Voltage range is selected!
Note:
Maximum value is limited to 4000.
Ph. Curr. Clamps
Select phase clamps for phase current
measurements.
Note:
For Smart clamps (A 1227, A 1281) always
select “Smart clamps”.
Note:
See section 4.2.3 for details regarding further
clamps settings.
N. Curr. Clams
Select clamps for neutral current measurements.
Note:
For Smart clamps (A 1227, A 1281) always
select “Smart type clamps”.
Note:
See section 4.2.3 for details regarding further
clamps settings.