MD 9250 Clamp Meter Series
Specifications
16
5 Specifications
Display:
3-5/6 digits 6000 counts. & 3-1/2 digits 1,999
counts for Hz
Polarity:
Automatic
Update Rate:
5 per second nominal;
Operating Temperature:
0
C to 40
C
Relative Humidity:
Maximum relative humidity 80% for temperature up
to 31
C decreasing linearly to 50% relative
humidity at 40
C
Pollution degree:
2
Storage Temperature:
-20
C to 60
C, < 80% R.H. (with battery removed)
Altitude:
Operating below 2000m
Temperature Coefficient:
nominal 0.15 x (specified accuracy)/
C @(0
C --
18
C or 28
C -- 40
C), or otherwise specified
Sensing:
True RMS
Safety:
Double insulation per IEC/EN61010-1 2nd Ed.,
UL61010-1 2nd Ed. & CAN/CSA C22.2 No.
61010.1-0.92 to CAT IV 1000V AC & DC
Transient Protection:
12kV (1.2/50
s surge)
Overload Protections:
Clamp-on jaws:
2000A rms continuous
“ + “ & COM Terminals (all other functions): 1000V rms
E.M.C.:
Meets EN61326-1:2006 (EN55022, EN61000-3-2,
EN61000-3-3, EN61000-4-2, EN61000-4-3,
EN61000-4-4, , EN61000-4-5, EN61000-4-6,
EN61000-4-8, EN61000-4-11)
In an RF field of 3V/m:
Capacitance function is not specified
Other function ranges:
Total Accuracy = Specified Ac 200 digits
Performance above 3V/m is not specified
Power Supply:
1.5V AA Size (IEC LR6) battery X 2
Power Consumption:
Typical 14mA for Current functions, and 5.2mA for
others
Low Battery:
Below approx. 2.4V
APO Timing:
Idle for 34 minutes
APO Consumption:
10
A typical
Dimension:
L264mm X W97mm X H43mm
Weight:
608 gm
Jaw opening & Conductor diameter: 55mm max
Accessories:
Test leads (pair), user's manual, banana plug K-
type thermocouple x 1
Optional purchase accessories:
USB interface kit AMD 9250; AMD9024 banana
plug to type-K socket plug adaptor
Special Features:
AutoCheck
TM
V&
; VFD-V & VFD-Hz; Backlighted
LCD; 5ms CREST-MAX Capture mode (Peak
Hold); Auto-ranging Relative-Zero mode; Display
Hold; EF-Detection (NCV); Optional Interface
capabilities with PC computers