MI 3311 GammaGT 

Main menu and test modes 

 

22

 

4.8.2 Language selection 

The instrument language can be set in this menu. 

 

 

Figure 4.10: Language menu

 

Keys: 

 

 / 

 

Selects the language. 

TEST 

Confirms selection and returns to 

Setup menu.

 

ESC 

Returns to 

Setup menu

 without changes. 

 

4.8.3 LCD contrast and backlight 

In this menu the contrast and backlight mode of the LCD can be set. 

 

 

Figure 4.11: LCD menu

 

 

Backlight modes: 

 

AUTO 

The high backlight level is active for 30 seconds after last pressing of any key. 
Then the backlight level returns to low level until a key is pressed again. 

OFF 

Backlight level is low. 

ON 

Backlight level is high. 

 

Keys: 

 

TAB 

Toggles between setup of contrast and backlight 

 / 

 

Sets contrast value or  backlight mode 

TEST 

Confirms selection and returns to 

Setup menu.

 

ESC 

Returns to 

Setup menu

 without changes. 

 

Note: 

  If you press the down (

) key while starting up the instrument you will automatically 

jump to the LCD contrast menu.   

 

4.8.4  Checkbox 

The in-built Checkbox provides a simple and effective means of checking the calibration of 
the GammaGT instrument and accessories. According to the Code of Practice the on-
going accuracy of the PAT tester should be verified at regular intervals and recorded. This 
is of special importance if the PAT tester is used on a daily basis. The GammaGT includes 
an in-built Calibration Unit (‘Checkbox’) that is independent from the other instrument’s 
electronic circuitry and connected directly to the output terminals of the instrument. During 
the calibration with the in-built Checkbox all main instrument functions and accessories 
can be verified. The calibration results are automatically stored into the instrument’s 
memory and can be viewed with the PAT Link PRO PC software. 

Содержание GammaGT MI 3311

Страница 1: ...GammaGT MI 3311 Instruction manual Ver 1 8 Code no 20 751 625...

Страница 2: ...el si 2010 Metrel Mark on your equipment certifies that this equipment meets the requirements of the EU European Union concerning safety and electromagnetic compatibility regulations No part of this p...

Страница 3: ...Checkbox specification 17 3 7 General data 17 4 Main menu and test modes 19 4 1 Instrument Main menu 19 4 2 VDE Organizer menu 19 4 3 Single test menu 20 4 4 Custom Autotest menu 20 4 5 Simple test me...

Страница 4: ...te leakage measurement 47 6 5 5 Insulation resistance P measurement 48 6 5 6 Substitute Leakage P measurement 48 6 5 7 Polarity test 49 6 5 8 Functional test 49 6 6 Handling autotest results 50 7 Work...

Страница 5: ...results parameters approx 1500 tests can be stored two communication ports USB and RS232C for communication with PC barcode reader printer and RFID reader writer built in real time clock fully compat...

Страница 6: ...off the instrument before opening the battery compartment Instrument contains rechargeable NiCd or NiMh battery cells The cells should only be replaced with the same type as defined on the battery pl...

Страница 7: ...igure 1 2 Power supply socket polarity Symbols Indication of battery charging Figure 1 3 Charging indication on display Before opening the battery compartment cover disconnect all accessories connecte...

Страница 8: ...e batteries to their normal capacity Note The charger in the instrument is a pack cell charger This means that the battery cells are connected in series during the charging The battery cells have to b...

Страница 9: ...ombined measuring equipment for testing measuring and monitoring of protective measures VDE 0404 1 Testing and measuring equipment for checking the electric safety of electric devices Part 1 General r...

Страница 10: ...g confirms selected option 5 UP 6 DOWN Selects parameter changes value of selected parameter 7 MEM Store recall clear tests in memory of instrument 8 TAB Selects the parameters item option in selected...

Страница 11: ...ion of external charger 5 USB connector Communication with PC USB 1 1 port 6 PS 2 connector Communication with barcode reader Communication with printer Communication with RFID reader writer Communica...

Страница 12: ...erts for side belt 2 Battery compartment cover 3 Fixing screw for battery compartment cover 4 Back side information label 5 Holder for inclined position of the instrument Figure 2 4 Battery compartmen...

Страница 13: ...power consumption or it is not connected switched off contains a fuse that has blown Select PROCEED or CANCEL WARNING Voltage on mains test socket between LN PE terminals is higher than approximately...

Страница 14: ...e mains cable of appliance during the test Check that the device under test is switched on to ensure that the complete circuit is tested Connect the cord to be tested to the IEC test terminal Test pas...

Страница 15: ...e measurement Test terminals Earth continuity PE mains test socket S EB1 probe 3 2 Insulation resistance Range Resolution Accuracy 0 00 M 19 99 M 0 01 M 20 0 M 49 9 M 0 1 M 5 of reading 3 digits 50 0...

Страница 16: ...A 0 75 mA 1 00 mA 1 50 mA 2 00 mA mA Test duration 2 s 3 s 5 s 10 s 30 s 60 s 120 s Displayed current calculated to appliance nominal mains supply voltage 230 V x 1 06 Test terminals Sub leakage LN ma...

Страница 17: ...attery or accu size AA Operation typical 8 h Charger socket input voltage 12 V 10 Charger socket input current 400 mA max Battery charging current 250 mA internally regulated Overvoltage category 300...

Страница 18: ...manual for particular function Communication transfer speed RS232 interface 9600 bps 1 start bit 8 data bits 1 stop bit RS232 connector PS 2 connector female USB interface 115200 bps USB connector ty...

Страница 19: ...tom autotest SIMPLE TEST simple pre programmed sequences see chapter 6 3 Simple test CODE AUTOTEST code based test sequences suitable for working with barcodes and RFID tags see chapter 6 4 Code Autot...

Страница 20: ...test can also be downloaded to and or uploaded from the PC SW PATLink PRO Plus Figure 4 4 Custom Autotest menu See chapter 6 2 Custom autotest for detailed description about this test mode 4 5 Simple...

Страница 21: ...MEMORY to recall print or clear stored results print labels and write RFID tags LANGUAGE instrument language LCD LCD contrast and backlight settings CHECKBOX to access the internal calibration functi...

Страница 22: ...etup menu ESC Returns to Setup menu without changes Note If you press the down key while starting up the instrument you will automatically jump to the LCD contrast menu 4 8 4 Checkbox The in built Che...

Страница 23: ...calibration procedure Switches between Checkbox screens ESC Returns to Setup menu without changes Carrying out the instrument calibration The Checkbox instrument calibration starting screen is display...

Страница 24: ...cord check After all steps were carried out the measured values together with an overall indication are displayed in the CAL column Figure 4 16 Examples of Checkbox result screens Meaning of indicatio...

Страница 25: ...e When enabling the fast mode then Visual Inspection and Functional Test will automatically be set to PASS 4 8 6 Setting date and time Date and time can be set in this menu Figure 4 18 Date and time m...

Страница 26: ...enu ESC Returns to Setup menu without changes TAB Enters Edit user data menu Edit user data Figure 4 20 Edit user data menu Keys Selects a letter TEST Selects the next letter MEM Confirms name and ret...

Страница 27: ...Instrument data screens TEST ESC Returns to Setup menu 4 8 9 Initial settings In this menu the following instrument parameters can be set to their initial values all measurement parameters in single t...

Страница 28: ...ngle test Measuring menu ESC Returns to Main menu A single test can be started from any Single test measuring menu Before carrying out a test the parameters limits can be edited Figure 5 2 Example of...

Страница 29: ...rs protection covers and barriers must be checked Are there signs of corrosion Are there signs of overheating Inscriptions and markings related to safety must be clearly readable Installation of the d...

Страница 30: ...for Earth continuity resistance measurement LIMIT Maximum resistance 0 10 0 20 0 30 0 40 0 50 0 60 0 70 0 80 0 90 1 00 TIME Measuring time 2 s 5 s 10 s 30 s 60 s 120 s s Test circuits for earth contin...

Страница 31: ...Insulation menu Test parameters for insulation resistance measurement OUTPUT Test voltage 250 V 500 V LIMIT Minimum resistance 0 01 M 0 10 M 0 25 M 0 30 M 0 50 M 1 M 2 M 4 M 7 M 10 M M TIME Measuring...

Страница 32: ...ation material etc The instrument measures the insulation resistance between Main test socket L N and S EB1 test terminal This function is primarily intended for testing Class II appliances and Class...

Страница 33: ...screws handles etc are checked with this test Capacitive leakage paths are included in the result too The test measures the current flowing at a test voltage of 30 VAC and the result is scaled to the...

Страница 34: ...th leakage measurements will be affected by the presence of filter capacitors between neutral and earth 5 2 6 Substitute leakage P Leakage currents between live conductors and isolated accessible meta...

Страница 35: ...rs Connect device under test to the instrument see figure 5 17 Press the TEST key for measurement Store the result by pressing MEM key optional Figure 5 18 Example of substitute leakage P current meas...

Страница 36: ...rcuit for polarity test Figure 5 20 Polarity test of IEC cord Polarity test procedure Select the POLARITY function Connect the IEC cord to the instrument as shown on figure 5 20 Press the TEST key for...

Страница 37: ...s operating RCDs and other disconnection devices How hot the appliance becomes during operation Rotating parts fans etc Power consumption Lamps and indicators Etc Especially safety relevant functions...

Страница 38: ...ument and into the normal mains supply as shown on figure 5 23 Store the result by pressing MEM key optional Figure 5 24 Voltage TRMS test result Note One of the safety features incorporated into the...

Страница 39: ...if the alternative tests are applicable Refer to chapter 1 1 Warnings for more information 6 1 VDE organizer setup menu Select VDE Organizer in Main menu In first step the appliance type means of pro...

Страница 40: ...ally goes to the Result menu Visual inspection Measurement is described in chapter 5 2 1 Visual inspection Figure 6 2 VDE Organizer Visual inspection menu Options in Visual test PASS FAIL To be applie...

Страница 41: ...result screen There are no special options to be set in the insulation resistance result screen Substitute leakage measurement This test is offered if it is applicable according to the VDE Organizer s...

Страница 42: ...tance P result screen NEXT Proceeds with the next measurement REPEAT Repeats the test use in case of multiple isolated SELV PELV accessible points Lowest result will be stored Substitute Leakage P cur...

Страница 43: ...unctional test Figure 6 13 VDE Organizer Functional result screens Options in Functional test result screen PASS FAIL To be applied manually Option if IEC leads multiple outlets without electronic par...

Страница 44: ...Custom Autotest function in Main menu Figure 6 14 Custom autotest menu Keys Selects the custom autotest START Starts the selected custom autotest See chapter 6 5 Carrying out Code Simple and Custom a...

Страница 45: ...autotest sequence with barcode reader RFID reader writer or manually Connect barcode reader or RFID reader writer to the instrument RS232 PS2 connector first Figure 6 16 Connection of the barcode rea...

Страница 46: ...ed or if any test fails the test sequence is stopped and the instrument automatically goes to the Result menu If a test parameter limit duration output voltage is changed the setup is valid only for t...

Страница 47: ...g screen After the measurement is carried out the Insulation result screen is displayed Figure 6 22 Insulation result screen There are no special options to be set in the Insulation resistance result...

Страница 48: ...next measurement REPEAT Repeats the test use in case of multiple isolated SELV PELV accessible points Lowest result will be stored 6 5 6 Substitute Leakage P measurement The test is offered if it is a...

Страница 49: ...measurement is carried out the Polarity test result screen is displayed Figure 6 30 Polarity test result screen There are no special options in the Polarity test starting screen 6 5 8 Functional test...

Страница 50: ...s ESC Returns to Code Simple or Custom menu Viewing autotest results In the View results screen performed tests results and their PASS FAIL status are displayed Furthermore the selected test results c...

Страница 51: ...last character of Appliance ID ESC CANCEL Returns to the previous menu An Appliance ID of up to 14 numeric characters can be entered The Appliance ID can also be scanned with a barcode reader or RFID...

Страница 52: ...e and time of the selected test The overall status of the selected test Figure 7 5 Recall results menu Keys TEST Enters View results menu for viewing autotest results ESC Returns to Memory menu Note M...

Страница 53: ...w of the display the following data is displayed Appliance ID NAME date and time of the selected test the overall status of the selected test Figure 7 8 Delete results menu Keys TEST Deletes selected...

Страница 54: ...nt simple label Print label Print results and Write RFID tag Figure 7 12 Printer RFID menu Keys Selects action TEST Confirms and executes selected action ESC Exits back to Print data RFID menu without...

Страница 55: ...SB or RS 232 The instrument automatically selects the communication mode according to the detected interface USB interface has priority How to transfer stored data RS 232 communication connect a PC CO...

Страница 56: ...person only 9 2 Service For repairs under or out of warranty please contact your distributor for further information Unauthorized person is not allowed to open the GammaGT instrument There are no user...

Страница 57: ...t GammaGT Small soft carrying bag Test probe black Crocodile clip black Test lead 1 5 m black IEC cord 2 m 6 x AA NiMH batteries Power supply adapter Instruction manual Calibration certificate Optiona...

Страница 58: ...otest codes can also be represented by the barcode Using the barcode reader the instruments can accept autotest code from barcode label A0 1 Autotest code Also appliance ID can be read from barcode la...

Страница 59: ...y Time 5 s 5 s 5 s 5 s Output 500 V 500 V Limit 1 00 M 1 00 M Insulation Time 5 s 5 s Output 500 V Limit 2 00 M Insulation probe Time 5 s Output 40 V 40 V Limit 3 50 mA 3 50 mA Sub leakage Time 5 s 5...

Страница 60: ...V 500 V Limit 1 00 M 1 00 M Insulation Time 5 s 5 s Output 500 V 500 V Limit 2 00 M 2 00 M Insulation probe Time 5 s 5 s Output Limit Sub leakage Time Output 40 V Limit 0 50 mA Sub leakage probe Time...

Страница 61: ...Limit Earth continuity Time Output Limit Insulation Time Output 500 V 500 V Limit 2 00 M 0 250 M Insulation probe Time 5 s 5 s Output Limit Sub leakage Time Output Limit Sub leakage probe Time Output...

Страница 62: ...er hrungsstrom werden eingestellt Earth bond 0 30 Leakage 3 50 mA Touch leakage 0 50 mA A0 4 A05 Kl_2_Iso Testing according to VDE Class 2 device with isolated accessible conductive parts Insulation r...

Страница 63: ...sting according to VDE Class 2 device without any isolated accessible conductive parts A1 0 A11 Kl_3_Iso Testing according to VDE Class 3 device with isolated accessible conductive parts Insulation P...

Страница 64: ...mA 200 mA Limit 0 30 0 30 0 30 Earth continuity Time 5 s 5 s 5 s Output 500 V 500 V Limit 1 00 M 0 50 M Insulation Time 5 s 5 s Output 500 V Limit 2 00 M Insulation probe Time 5 s Output 40 V Limit 7...

Страница 65: ...0 0 30 0 50 Earth continuity Time 5 s 5 s 5 s Output 500 V 500 V 500 V Limit 1 00 M 1 00 M 1 00 M Insulation Time 5 s 5 s 5 s Output 500 V 500 V Limit 0 50 M 2 00 M Insulation probe Time 5 s 5 s Outpu...

Страница 66: ...Output 200 mA 200 mA Limit 0 70 1 00 Earth continuity Time 5 s 5 s Output 500 V 500 V Limit 1 00 M 1 00 M Insulation Time 5 s 5 s Output Limit Insulation probe Time Output Limit Sub leakage Time Outp...

Страница 67: ...ty Time 5 s Output 500 V Limit 1 00 M Insulation Time 2 s Output 500 V 500 V Limit 2 00 M 0 25 M Insulation probe Time 2 s 5 s Output 40 V Limit 3 50 mA Sub leakage Time 2 s Output 40 V Limit 0 50 mA...

Страница 68: ...MI 3311 GammaGT Appendix D Simple test codes 68...