A 1472, A1579                                                 Leakage current clamp 

 

20 

 

 

 

 

Fig. 11: Minimizing influencing quantities,  correct placement of conductors

 

 

 

 

 

 

 

Fig. 12: Minimizing influencing quantities, avoiding external magnetic and 

electric fields 

 

4.4.6  Consideration of influencing quantities 

 

Only high quality current clamp can accurately measure in range of mA in 
presence of external fields and high load currents. The given resolution or 
basic  accuracy  are  not  telling  a  lot  about  the  real  quality  of  the  current 
clamp. The immunity of the current clamp against  influencing quantities is 
much more representative. 
The  IEC  (International  Electrotechnical  Commission)  is  preparing  a  new 
international  standard 

IEC  61557-13  Hand-held  and  hand-manipulated 

current  clamps  and  sensors  for  measurement  of  leakage  currents  in 
electrical  distribution  systems

  that  will  cover  the  aspect  of  influencing 

quantities  and  will  be  launched  in  2010.  One  of  its  main  objectives  is  to 
define rules how to consider the immunity of influencing quantities in scope 
of  the  accuracy  of  leakage  current  clamps.  This  will  enable  the  user  to 
make  a  relevant  comparison  when  deciding  between  different  current 
clamp models. 

Содержание A1472

Страница 1: ...Leakage current clamp A 1472 A1579 User manual Version 1 2 Code no 20 752 271...

Страница 2: ...1 1 manual is valid also for A 1579 2014 2019 METREL Mark on your equipment certifies that this equipment meets the requirements of the EU European Union concerning safety and electromagnetic compatib...

Страница 3: ...mps 13 4 2 1 Direct method 13 4 2 2 Differential method 14 4 3 Testing procedures 15 4 3 1 Finding sources of excessive leakage current 15 4 3 2 Insulation resistance testing 17 4 4 Influencing quanti...

Страница 4: ...asuring instrument Consider all generally known precautions in order to avoid risk of electric shock while dealing with electric installations and equipment Do not extend hands over protection barrier...

Страница 5: ...uitable to work with the A1472 current clamp Main clamp parts 1 Current sensor 2 Protection barriers 3 Conductor opening 4 Signal output terminals 1 2 3 4 A 1472 2 1 Difference to A 1579 The leakage c...

Страница 6: ...is possible to use the current clamp on non insulated conductors Symbol on the current clamp indicates possible hazardous live conditions if the required safety measures are ignored Symbol indicates...

Страница 7: ...e current clamps and leave them to dry totally before using it Notes Do not use liquids based on petrol or hydrocarbons Do not spill cleaning liquid over the current clamps To clean jaw cut surfaces u...

Страница 8: ...rent and insulation resistance between line and earth is defined by the Ohm s law _ k R U mA I ins nom res leak Eq 1 Ileak_res resistive leakage current in mA Unom nominal line voltage e g 230 V Rins...

Страница 9: ...very high frequency In general leakage currents in normal operation increase with the size of the electrical installation and the number of connected loads Table 1 shows the values of leakage currents...

Страница 10: ...in the installation this can result in an improper current distribution through the neutral and PE conductors Typical faults that usually results in a high fault current that will immediately trip th...

Страница 11: ...eakage currents can have different non sinusoidal shapes They are caused by nonlinearity of loads nonlinearity of insulation materials and presence of harmonic voltages and currents in the power syste...

Страница 12: ...conductive parts In case of inadequate protective measures this voltage can increase above a safe level and presents a danger for the people if it would be touched Furthermore if the protection devic...

Страница 13: ...out R N I U 1 Eq 3 L N L out R N I I U 1 Eq 4 Uout output voltage of current sensor IPE IL IN measured currents RB load burden resistor N number of turns of current clamp The main advantage of the mea...

Страница 14: ...conductors earth the sum of currents through active conductors must be exactly zero regardless of the load currents If a leakage current flows it is of the same size as the measured sum between active...

Страница 15: ...ferent parallel paths Refer to METRELs handbook Guide for testing and verification of low voltage installations for more information about parallel paths 2 All suspicious loads should be connected If...

Страница 16: ...g any of the procedures described above the faulty point can be determined very precisely Fig 7 Finding source of excessive leakage current by disconnection Fig 8 Finding source of excessive leakage c...

Страница 17: ...he capacitance The applicability of leakage current measurement as an alternative to the insulation resistance test depends on following factors The ratio between capacitive and resistive leakage curr...

Страница 18: ...rrent sensor quality Protective measures are proper magnetic shielding symmetry of the winding twisting of all connection cables high permeability of the sensor core etc 4 4 2 Influence of load curren...

Страница 19: ...rby current carrying conductors switchboards etc If the results strongly fluctuate in regard to the current clamp position it is unlikely to get an usable result The reading is not zero although no cu...

Страница 20: ...p The immunity of the current clamp against influencing quantities is much more representative The IEC International Electrotechnical Commission is preparing a new international standard IEC 61557 13...

Страница 21: ...be beneficial in most of applications 4 5 3 Protection category Overvoltage protection Take care above the overvoltage category of the clamp For use in offices residential industrial domestic sites a...

Страница 22: ...ty banana sockets 4 mm Guard terminal Electrical characteristics Measuring range 0 5 mA 10 A Accuracy 5 Frequency range 40 Hz 5 kHz Max input current 100 A Safety specification Type A application arou...

Отзывы: