RAW
Count
VWC
(m
3
/m
3
)
METER logger
TEROS 11/12
2650 to
2730
0.332 to
0.363
TEROS 10
2650 to
2800
0.356 to
0.419
Non-METER logger
TEROS 10
1940 to
2050
mV
0.356 to
0.419
Installation
TEROS VERIFICATION CLIP QUICK START
Preparation
Confirm that the verification clip components are intact. The
verification clip helps verify proper function and accuracy for
TEROS 10, 11, and 12 soil moisture sensors prior to installation. A
data acquisition system and the appropriate software are required,
such as the
and
or ZENTRA Utility Mobile.
ATTENTION
For best results, use the latest versions of METER software and
firmware for the computer or mobile device, ZENTRA products,
and sensors. Please use the software Help menu to find updates.
Consult the sensor user manual for more troubleshooting tips.
18299-02
5.31.2020
1. Place Verification Clip
on Sensor
Slide the verification clip onto the
two prongs farthest away from
the sensor cable with the arrow
pointing toward the sensor.
WARNING
Risk of personal injury is present.
Sensor needle points are extremely
sharp and will puncture skin.
Please handle with care.
2. Ensure Correct Placement
The verification clip should
rest snugly against the
sensor body.
3. Connect TEROS Sensor
to Data Logger
Plug the sensor into the data
logger and use the
SCAN
function
in the software.
4. Verify Sensor Function
and Accuracy
The sensor should read within
the appropriate ranges.