27
Figure 11 Down spout
Figure 12 Teflon screen
Figure 13 Rain funnel
4.3 TROUBLESHOOTING
Table 4 Troubleshooting the ATMOS 41
Problem
Possible Solutions
ATMOS 41 not responding
Check power to the sensor.
Check sensor cable and 3.5-mm plug integrity.
Check data logger wiring to ensure the following connections:
• Brown—3.6 V to 15 V power supply
• Orange—digital out
• Bare—ground
If sensor does not respond, use the ProCheck to make sure it is
working satisfactorily.
Rain funnel
Teflon screen
Down spout