42
SERVICE
Figure 51 Capacitance sensor calibration screen
4. Highlight
Type
and select
ENTER to toggle to
Multipoint
.
5. Select
Start
.
Prompts will instruct to insert a fresh standard and seal the chamber (
).
Figure 52 Prompt to insert fresh standard
6. Empty the whole vial of 0.25
a
w
standard solution into a sample cup.
7. Ensure the rim and outside of the cup are clean.
8. Place the sample cup in the sample chamber.
9. Carefully close the lid and move the lever to the READ position.
10. Select
OK to begin testing.
11. After AQUALAB TDL has finished measuring the verification standard, a new screen
appears requesting that a 0.50
a
w
standard be placed in the chamber.
12. Repeat
steps
through
using 0.50, 0.76, and 0.92
a
w
standards.
NOTE: To end the multipoint calibration program, select CANCEL to return to the previous screen.
When measurements are complete on all four standards, the Verification Complete
screen appears showing the testing results for each standard (
).