USING APOGEE UV SENSORS WITH ZENTRA SYSTEM

3.  DATA INTERPRETATION

Apogee UV sensors used with the ZENTRA system report UV-A irradiance in energetic flux 
density units of watts per square meter (W/m

2

). Additionally, the sensor orientation information 

is provided in the metadata tab of ZENTRA Cloud and ZENTRA Utility file downloads 
(Microsoft

®

 Excel

®

 spreadsheet). Sensor orientation is reported as the zenith angle in units of 

degrees, with a zenith angle of 0° indicating a sensor oriented straight upwards.

4.  TROUBLESHOOTING

This troubleshooting section details possible major problems and their solutions. If the 
problem is not listed or these solutions do not solve the issue, contact 

Customer Support

.

Table 2  Troubleshooting

Problem

Possible Solution

Sensor not responding

Check power to the sensor and logger.

Check sensor cable and stereo plug connector integrity.

Check that the SDI-12 address of the sensor is 

0

 (factory default). 

Check this with ZENTRA Utility by going to Actions, select Digital 

sensor terminal, choose the port that the sensor is on, and send 

the 

?I!

 command to the sensor from the dropdown menu.

Sensor values are not reasonable Verify the sensor is not shaded.

Verify the orientation angle of sensors.

Cable or stereo plug 

connector failure

If the stereo plug connector is damaged or needs to be replaced, 

contac

Customer Support

 for a replacement connector or  

splice kit.

If a cable is damaged refer to the METER 

wire-splicing guide

 for 

cable repair.

It is recommended that Apogee UV sensors are returned for factory recalibration every 
2 years. Visit 

Apogee repairs

 (apogeeinstruments.com/recalibration-and-repairs) or contact 

Apogee Technical Support

 ([email protected]) for details.

Содержание APOGEE SU-221

Страница 1: ...APOGEE UV SENSOR...

Страница 2: ...CONTENTS 1 Introduction 2 2 Installation 2 2 1 Set Up Mounting Assembly 4 2 2 Connect to METER ZENTRA Series Logger 5 3 Data Interpretation 6 4 Troubleshooting 6 5 Customer Support 7 Index 8 18443 00...

Страница 3: ...1 p a g e e I N S T R U M E N T S Made in USA...

Страница 4: ...y before going out to the field For more information on Apogee sensors please review the Ultraviolet A Radiation Sensor User Manual or visit the Apogee UV sensor product page apogeeinstruments com uv...

Страница 5: ...Northern hemisphere or true South in the Southern hemisphere to reduce azimuth error Secure the System Tighten the U bolt nuts by hand until hand tight and then tighten with a wrench CAUTION Do not ov...

Страница 6: ...ensor M8 connector holes and seat connectors fully 2 Tighten the cable screw until hand tight Figure 1 M8 connectors are easy to overtighten Do not use pliers or other tools to tighten this connector...

Страница 7: ...data loggers 1 Plug the stereo plug connector into one of the sensor ports on the logger Figure 4 Port 1 Port 2 Port 3 Port 4 Port 5 Port 6 TEST OK ERROR Conduit Figure 4 Logger connection 2 Connect t...

Страница 8: ...not responding Check power to the sensor and logger Check sensor cable and stereo plug connector integrity Check that the SDI 12 address of the sensor is 0 factory default Check this with ZENTRA Util...

Страница 9: ...Customer support representatives are available for questions problems or feedback Monday through Friday 8 00 to 17 00 Central European time Email support europe metergroup com sales europe metergroup...

Страница 10: ...I installation mounting 3 4 preparation 3 tools needed 2 M mounting 3 mounting bracket 2 4 R recalibration 6 S stereo plug connector 5 T troubleshooting 6 U U bolt 3 4 ultraviolet radiation 2 4 user m...

Страница 11: ...ins Court Pullman WA 99163 T 1 509 332 2756 F 1 509 332 5158 E info metergroup com W metergroup com METER Group AG Mettlacher Stra e 8 81379 M nchen T 49 89 1266520 F 49 89 12665220 E info europe mete...

Страница 12: ......

Отзывы: