background image

4

USING APOGEE OXYGEN SENSORS WITH ZENTRA SYSTEM

Check the METER download webpage for the most recent data logger firmware. Logger 
configuration may be done using either ZENTRA Utility (desktop and mobile application) or 
ZENTRA Cloud (web-based application for cell-enabled ZENTRA data loggers).
1.  Plug the stereo plug connector into one of the sensor ports on the logger (

Figure 2

).

Port 1

Port 2

Port 3

Port 4 

Port 5

Port 6

TEST

OK

ERROR 

(-)

(-)

(-)

(-)

(-)

(+)

(-)

(+)

(+)

(+)

(+)

(+)

Conduit

Figure 2  Logger connection

2.  Connect to the data logger via ZENTRA Utility with a laptop and USB cable or ZENTRA 

Utility Mobile app with a mobile device supporting Bluetooth communication.

3.  Use ZENTRA Utility to scan the ports and make sure the sensors were properly identified 

by the logger and are reading properly.
METER data loggers should automatically recognize the Apogee sensor.

4.  Use ZENTRA Utility to set the measurement interval.
5.  Use ZENTRA Utility to configure communication settings for data transfer to ZENTRA Cloud.

Sensor data can be downloaded from METER data loggers using either ZENTRA Utility or 
ZENTRA Cloud. Refer to the logger user manual for more information.

2.2  SENSOR CALIBRATION

Before installing the oxygen sensor, it must be calibrated at ambient oxygen concentration 
(20.95%) at the elevation of the measurement site. Galvanic cell type oxygen sensors 
respond to the absolute oxygen concentration, which is dependent on absolute (barometric) 
pressure and temperature in addition to the oxygen concentration of air.  Therefore, oxygen 
sensors are not calibrated at the factory and must be calibrated by the user before first use.

It is recommended that Apogee oxygen sensors be recalibrated periodically to maintain best 
accuracy. Refer to the 

SO-411 user manual

 for additional information (apogeeinstruments.

com/so-411-sdi-12-soil-response-thermistor-reference-oxygen-sensor).

Apogee sensors can be calibrated with a METER ZSC Bluetooth device or ZL6 data logger.

Содержание Apogee SO-411

Страница 1: ...APOGEE OXYGEN SENSOR...

Страница 2: ...Introduction 2 2 Installation 2 2 1 Connect to METER ZENTRA Series Logger 3 2 2 Sensor Calibration 4 2 3 Heater Power Supply 7 3 Data Interpretation 7 4 Troubleshooting 7 5 Customer Support 8 Index 1...

Страница 3: ...ap gee INSTRUMENTS Made in USA...

Страница 4: ...Apogee SO 411 Oxygen Sensor preconfigured by METER included in order Apogee AO 001 Diffusion Head included in order Shovel or Auger if application requires sensor burial METER ZENTRA series data logg...

Страница 5: ...er and secure cables between the sensors and the data logger to the mounting post in one or more places to ensure cable weight does not pull the plug free from its port Connect to Data Logger Plug the...

Страница 6: ...measurement interval 5 Use ZENTRA Utility to configure communication settings for data transfer to ZENTRA Cloud Sensor data can be downloaded from METER data loggers using either ZENTRA Utility or ZEN...

Страница 7: ...A Utility Mobile on an iOS or Android device 3 Connect to the ZSC Bluetooth device 4 Check that the SO 411 is communicating correctly with the ZSC an inaccurate oxygen concentration and a temperature...

Страница 8: ...ygen concentration now reads very close to 20 95 If Apogee oxygen sensor calibration is not listed under Device Sensor Tools the calibration can be done using a terminal window 1 Connect the Apogee ox...

Страница 9: ...be a simple wall wart If mains power is unavailable a 12 V battery is appropriate and a solar panel and charge controller can be used to maintain battery charge Under most conditions a small battery 7...

Страница 10: ...ted periodically to maintain best accuracy See Section 2 2 for directions on calibrating the sensor Refer to the SO 411 User Manual apogeeinstruments com so 411 sdi 12 soil response thermistor referen...

Страница 11: ...by email please include the following information Name Address Phone Email address Instrument serial number Description of the problem NOTE For products purchased through a distributor please contact...

Страница 12: ...ies data loggers I installation 2 3 above ground 3 buried 3 preparation 2 tools needed 2 internal heater 2 3 7 O oxygen concentration 2 4 oxygen sensor 2 S stereo plug connector 3 T troubleshooting 7...

Страница 13: ...ins Court Pullman WA 99163 T 1 509 332 2756 F 1 509 332 5158 E info metergroup com W metergroup com METER Group AG Mettlacher Stra e 8 81379 M nchen T 49 89 1266520 F 49 89 12665220 E info europe mete...

Страница 14: ......

Отзывы: