![METER Group, Inc. Spectral Reflectance Sensor Скачать руководство пользователя страница 28](http://html1.mh-extra.com/html/meter-group-inc/spectral-reflectance-sensor/spectral-reflectance-sensor_operators-manual_1777701028.webp)
5
CONNECTING THE SRS
SRS Sensors
wish to connect it to a non-METER data logger, use one of the fol-
lowing two options.
Option 1
1. Clip off the stereo plug connector on the sensor cable
2. Strip and tin the wires.
3. Wire it directly into the data logger.
This option has the advantage of creating a direct connection and
minimizes the chance of the sensor becoming unplugged. However,
it then cannot be easily used in the future with a METER readout
unit or data logger.
Option 2
Obtain an adapter cable from METER.
The adapter cable has a connector for the stereo plug connector on
one end and three wires (or pigtail adapter) for connection to a data
logger. The stripped and tinned adapter cable wires have the same
termination as in Figure 10; the brown wire is excitation, the orange
is output, and the bare wire is ground.
Note: Be extra careful to secure your stereo to pigtail adapter con-
nections to ensure that sensors do not become disconnected during
use.
24