EX1200A Web Page Operation
125
o
Delay
: Indicates the time, in seconds, the DMM will wait after getting a trigger input
event before proceeding to the Sample Trigger layer. Valid values are -1.0 and 0 to
4294.967295. When set to -1.0 this is equivalent to using
Auto Delay
.
Sample Trigger
: Specifies the sample trigger event that cause the DMM to take additional
measurements. For every sample trigger event, the DMM takes one measurement.
o
Source
: Specifies the event type that will cause a sample trigger event. These include
Immediate
,
External
,
Software
,
Interval
, and the DMM backplane lines (
BPL0 – 7
).
o
Route
:
This button appears when a backplane line (
BPL0 – 7
) is selected for
Source
.
Clicking this button launches a dialog box that allows the user to select an external
trigger line (
LXI0 – 7, DIO0 – 7, LAN0 – 7,
or
ALARM0
) to connect to the chosen
backplane line.
o
Slope
: For hardware trigger lines, this parameter indicates the slope of level change,
either
Rising
or
Falling
, that will cause a sample trigger event.
o
Count
: Indicates the number of sample triggers that will be accepted before returning to
the Trigger layer. Valid values are -1 and 1 to 2,147,483,647, where “-1” sets an infinite
count.
o
Delay
: Indicates the time, in seconds, the DMM will wait after getting a sample trigger
event before it performs a measurement. Valid values are 0.0 to 4294.967295.
o
Interval
: When
Interval
is selected as the
Source
, this indicates the time, in seconds,
between sample trigger events. Valid values are 0.0 to 4294.967295.
Trigger Output
: These parameters specify the trigger events generated by the DMM.
o
Measurement Complete
: Upon completion of a measurement, the selected event will
occur. These events include
None
,
External
, and
BPL0 – 7
.
o
Over Limit
: When a measurements exceeds the
High
and/or
Low Limit
settings, the
selected trigger event is asserted. These include
None
,
External
, and
BPL0 – 7
.
o
Route
:
This button appears when a backplane line (
BPL0 – 7
) is selected for
Source
.
Clicking this button launches a dialog box that allows the user to select an external
trigger line (
LXI0 – 7, DIO0 – 7, LAN0 – 7,
or
ALARM0
) to connect to the chosen
backplane line.
o
High Limit
: Indicates the upper bound of the
Over Limit
event.
o
Low Limit
: Indicates the lower bound of the
Over Limit
event.
Notes on Web Page Use
If a measurement has been initiated and seems to be stuck in the
Waiting For Trigger
state, check
the
Trigger Properties
field. In order for a scan to begin immediately without interaction with
other instrumentation, both inputs (the
Trigger Input Source
and
Sample Trigger Source
) should
be set to
Immediate
.
If the EX1200A stops taking measurements unexpectedly, check the
Trigger Input Count
and
Sample Trigger Count
settings. They control the number of measurements taken before the
DMM needs to be initiated again.
Advanced Settings
Math
: Allows the user to make corrections to the measurement readings using an
m
x +
b
equation.
o
Gain
: Sets the gain correction factor. Represents the
m
in the
m
x + b equation. The
default value is gain = 1.
o
Offset
: Sets the offset correction factor. Represents the
b
in the mx +
b
equation. The
default value is gain = 0.
NOTE
Incorrectly setting the gain and offset parameters can result in invalid measurements.
Содержание EX1200A SERIES
Страница 16: ...VTI Instruments Corp 16 EX1200A Series Introduction EX1200A SERIES 1 U FULL RACK MAINFRAME DIMENSIONAL DIAGRAM...
Страница 17: ...www vtiinstruments com EX1200A Series Introduction 17 EX1200A SERIES 3 U MAINFRAME DIMENSIONAL DIAGRAM...
Страница 32: ......
Страница 50: ...VTI Instruments Corp 50 EX1200A Series Preparation for Use...
Страница 63: ...www vtiinstruments com EX1200A Series Preparation for Use 63...
Страница 64: ......
Страница 72: ......
Страница 106: ...VTI Instruments Corp 106 EX1200A Triggering...
Страница 108: ...VTI Instruments Corp 108 EX1200A Triggering FIGURE 6 1 EX1200A SERIES TRIGGER MODEL...
Страница 142: ...VTI Instruments Corp 142 EX1200A Web Page Operation...
Страница 144: ...VTI Instruments Corp 144 EX1200A Series Programming...
Страница 154: ......
Страница 160: ......