![Met One E-BAM Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/met-one/e-bam/e-bam_operation-manual_1775877012.webp)
E-BAM-9805 Manual Rev B.docx
Page 12
5.
Optional Wind Sensor:
If an optional wind speed/direction sensor (such as the EX2-034B or
EX2-AIO) is purchased a cross-arm tube to use for mounting the wind sensor will be
included. Connect the cross-
arm to the post on the back of the tripod with the supplied ¾” x
¾” fitting and set screws and then install the sensor at end of the arm as suggested in Figure
2-4. The wind sensor should be as far from the E-BAM unit as possible, and the wind vane
must be able to rotate fully without hitting anything. Plug the wind sensor into the
corresponding connector on the bottom of the E-BAM. Depending on the type of wind sensor,
it may need to be oriented to the south. Consult the separate manual that comes with the
wind sensor for details. Verify configuration after power-up as detailed in section 2.7.2.
Figure 2-4 Sensor Mounting with an Optional Wind Sensor
Содержание E-BAM