Use the up and down arrow buttons to move the cursor to 

the desired setting and press 

MENU/SELECT

.  Adjust the 

following settings: 

 

CLOCK –  

Set to

 GMT time 

AVERAGE PERIOD –  Set to 60 MIN. 

CONCENTRATION

 

 

Set to AUTO; DAC to 1.0 V; K-FAC 

to 1.00. 

SAMPLING MODE – Set to CONTINUOUS. 

ALARM CONTACT – Does not apply. 

RH HEATER CONTROL – Set RH SETPOINT to 50%. 

ZERO MODE –  Set ZERO MODE to AUTO; ZERO 

PERIOD to 24 HR. 

SPAN MODE –  Set SPAN MODE to AUTO; SPAN 

PERIOD to 24 HR; RESTORE DEFAULT to 

NO. 

BAUD RATE –  

Set

 BAUD RATE to 9600. 

STATION ID –  Does not apply. 

ENGR UNITS –  Set to METRIC. 

 

Press the 

ESC 

button twice to get to the main screen. 

 

STARTING A RUN 

 

 

From the main screen, press the 

white button

 under TURN 

ON. 

 

 

 

Press the 

white button

 under TURN ON. 

 

 

The E-Sampler will go through a ZERO and SPAN check and 

then begin to sample.  After several minutes the screen will go 

blank to conserve power.  Press the 

ESC

 button to turn the 

display back on.   

 

STARTING THE SATELLITE TELEMETRY 

SYSTEM AND VERIFYING COMMUNICATIONS 

 

 

Open the telemetry system and turn the system on. 

 

The left light will illuminate red. 

 

The following diagram shows the various states of the 

telemetry system and corresponds to the LED indicators 

located inside the telemetry system.  LED indicators verify 

that the telemetry system has power, the modem is receiving 

information from the E-Sampler, and the antenna on top of 

the telemetry system is “seeing” satellites. 

 

 

 

 

 

 

 

 

 

A.

 

 Modem initialized light is on. 

B.

 

 Serial Data measurement Digital 0 from E-Sampler light is 

on. 

C.

 

 Serial Data measurement Digital 1 from E-Sampler light is 

on. 

D.

 

 Light is ON when satellite is in view and OFF between 

satellite passes.  (If this green LED light does not turn on 

within 30 minutes, move the telemetry system to a higher 

unobstructed position.)

 

 

 

STOPPING THE E-SAMPLER 

 

 

To stop the E-Sampler, press the 

ESC

 button to turn the 

display back on.   

 

Press the 

white

 

button

 under TURN OFF on the next two 

screens. 

 

 

 

TRANSFERRING DATA FROM THE E-SAMPLER 

TO A LAPTOP USING HYPERTERMINAL 

 

 

Disconnect the telemetry system serial communication cable 

from the E-Sampler. 

 

Connect the Met One serial communication cable to the 

appropriate connector on the bottom of the E-Sampler and 

to the serial connector on the laptop. 

 

On the laptop, click the HyperTerminal icon.  “Connected” 

should display on the bottom of the screen. 

 

On the top menu bar, click “Transfer>Capture Text.” 

 

On the laptop, press the return key several times.   The “>” 

symbol should appear on the screen. 

 

Press the number “2” button.   Data should begin to scroll 

across the screen. 

 

When data stops scrolling, click on “Transfer > Capture 

Text > Stop.” 

 

Close the HyperTerminal program. 

 

Remove the gel cell battery before 
shipping or moving the E-Sampler.   

 

WEB SITES AND PHONE NUMBERS 

 

 Missoula T & D Center 

 

http://www.fs.fed.us

/t-d                               

(Username: t-d, Password: t-d) 

 Phone: Andy Trent  406–329–3912 

Met One Instruments, Inc. 

http://www.metone.com

 

Phone:  541–471–7111 

     Airsis Telemetry System 

     

http://www.airsis.com

 

     Phone: Jim Drewett  858–586–0933

             Mary Ann Davies 406-329-3981

>>>>>>CAUTION<<<<<< 
  ALTER E-SAMPLER 
    OPERATION 
 CANCEL      TURN OFF 

08-JUL-2001 08:23:41 
  CONC:   0.022 MG/M3 
  FLOW:   2.0   LPM 
 TURN OFF 

>>>>>>CAUTION<<<<<< 
  ALTER E-SAMPLER 
    OPERATION 
 CANCEL       TURN ON 

08-JUL-2001 08:23:41 
  CONC:    UNIT OFF 
  FLOW:    0.0LPM 
 TURN ON 

Power Switch