AEROCET 532 Manual Rev B
Page 25
5.5.
About Screen
Figure 27 shows the
ABOUT
screen. The
ABOUT
screen shows
the manufacturer’s serial
number, the firmware version, Met One Instruments, Inc. service contact details, date of
last calibration, and instrument run time.
Figure 27
– About Screen
6.
Charging the Battery
Caution:
The provided battery charger is designed to work safely with this device. Do
not attempt to connect any other charger or adapter to this device. Doing so
may result in equipment damage.
To charge the battery, connect the battery charger to an AC power outlet and the DC
barrel connector to the socket on the left side of the AEROCET 532. The battery
charger is universal and will work with power line voltages of 100 to 240 volts, 50 to 60
Hz. A discharged battery pack will take approximately 2.5 hours to fully charge.
When fully charged the battery inside the AEROCET 532 will power the unit for over 8
hours of continuous sampling. For continuous operation, operate the unit with the
battery charger attached. Charge the battery before storing the AEROCET 532. Storing
a discharged battery will degrade its performance.
The Aerocet 532 can also be charged using the optional 83529 Docking Station. The
device will fit securely onto the Docking Station when encased in its rubber boot. If the
rubber boot is not used, an 83584 bootless adapter is placed at the base of the docking
station to hold the device.