107
Irinos IR Users Manual © 2015 - 2016 Messtechnik Sachs GmbH, Germany
Mounting & Cabling
o
Each I/O-Box must be addressed manually. Therefore the address
must be selected via decode switch or dip switch.
Valid addresses are 2, 4, 6, 8. Each address must only be unique.
o
The maximum length of the IO-Bus cabling is 20m.
Procedure
o
Select the addresses at the I/O-Boxes (->
or -
>
).
o
If I/O-Boxes of the hat-/din-rail version are used: activate the
termination via the DIP switch at the last box of the IO-Bus
cabling.
o
Connecting neighbouring I/O-Boxes using the
.
6.4.4
Connecting inductive probes
Connect the inductive probes to the M16 connectors. Please note:
o
The inductive probe type must fit the type of the Irinos-Box
.
Connect measurement probes "Tesa halfbridge or compatible"
only to an Irinos-Box IR-TFV-8-
TESA
-...
Connect measurement probes "Knäbel IET" only to an Irinos-Box
IR-TFV-8-
IET
-...
o
Make sure to lock all connectors. This ensures longer connector
lifetime.
o
Use protection caps for unused input channels. These are available
separately.
6.4.5
Connecting incremental encoders
Connect the incremental encoders (1Vpp or TTL/RS422) to the
DSUB receptables. Please note:
The same connector type is used for the
and the
.
The pinning of the ILink connector has been chosen to avoid
any damage of incremental encoders with standard pinning,
if they are plugged into the ILink interface. For encoders
with special pinning, this may be different.
Only use the appropriate connectors for the incremental
encoders.
59
60
37
46
66
88
Содержание Irinos IR
Страница 2: ......
Страница 7: ...Introduction...
Страница 15: ...Safety instructions...
Страница 22: ......
Страница 23: ...System overview...
Страница 39: ...Product descriptions...
Страница 64: ......
Страница 65: ...Pin assignment...
Страница 92: ......
Страница 93: ...Mounting Cabling...
Страница 109: ...Setup...
Страница 118: ...118 Irinos IR Users Manual 2015 2016 Messtechnik Sachs GmbH Germany Setup...
Страница 120: ......
Страница 121: ...Measurement Control via MscDll...
Страница 168: ......
Страница 169: ...Troubleshooting and first aid...
Страница 186: ...186 Irinos IR Users Manual 2015 2016 Messtechnik Sachs GmbH Germany Troubleshooting and first aid...
Страница 187: ...Maintenance cleaning and disposal...
Страница 190: ......
Страница 191: ...Application notes...
Страница 198: ......
Страница 199: ...Specifications dimensions...
Страница 209: ...209 Irinos IR Users Manual 2015 2016 Messtechnik Sachs GmbH Germany Specifications dimensions...
Страница 210: ...210 Irinos IR Users Manual 2015 2016 Messtechnik Sachs GmbH Germany Specifications dimensions...
Страница 211: ...211 Irinos IR Users Manual 2015 2016 Messtechnik Sachs GmbH Germany Specifications dimensions...
Страница 212: ...212 Irinos IR Users Manual 2015 2016 Messtechnik Sachs GmbH Germany Specifications dimensions...