Merits Health Products Pioneer Q150 Скачать руководство пользователя страница 6

 

5

3. PRECAUTIONS 

Always plug in a 15 AMP, grounded circuit. DO NOT USE AN EXTENSION CORD 
Place the unit where sides and bottom are unobstructed to ensure proper airflow. (Be sure that the 
machine is placed away from drapes or other objects that might block the flow of incoming air 
and/or alarm indicators.) 
Only operate the concentrator in a upright position. 
Do not use oxygen tubing that exceeds 50 feet in length.    The manufacturer recommends crush 
resistant tubing. 
Do not place liquids on the unit or use liquids to clean the unit.    Do not use petroleum-based 
solvents. Clean unit and power cord with mild household cleaner applied with a damp cloth or 
sponge, then wipe dry. 

The Salter Labs 7100 350c.c. Bubble Humidifier and the B&F Medical Bubble Humidifier are 
suitable for use with the Oxygen Concentrators. Or a humidifier capable of 30 Psi operation 
and 15 LPM of flow is required. Use of certain humidifiers not specified for use with this 
oxygen concentrator may impair the performance. 

Always choose the FDA or CE approved accessories, such as Masks or cannulas & 
Humidifiers to ensure the reliability and performance. 
Different personnel to prevent cross infection should never use the same mask or cannula. 

 
 
 

4. INTENDED USE 

The Merit Oxygen Concentrator is intended to be used to supply supplemental oxygen to patients in 

their home.    The devices are not intended for life support nor do they provide any patient monitoring 
capabilities. 

 
 

Содержание Pioneer Q150

Страница 1: ...PIONEER 0 1 1 1 2 3 4 5 6 7 5 1 1 8 9 2 7 Merits Health Products Co Ltd Q150 OXYGEN CONCENTRATOR OWNER S MANUAL...

Страница 2: ...1 Read this manual before using concentrator and save for future reference...

Страница 3: ...ation 10 8 2 Plug in Power Cord 10 8 3 Connect Humidifier if prescribed by your physician 10 8 4 Connect Oxygen Tubing 11 8 5 Turn On Power Switch 12 8 6 Adjust Flowrate 12 8 7 Oxygen Purity Indicator...

Страница 4: ...nts Alternating Current Type BF Equipment No Smoking or Naked Flames Use No Oil or Grease 1 CAUTIONS Federal Law restricts this device for sale or use on the order of a physician Do not adjust your pr...

Страница 5: ...vailable to avoid any interruption in supply This unit is not intended for life sustaining or life supporting applications Turn the unit off when it is not in use In the event of an alarm contact your...

Страница 6: ...er applied with a damp cloth or sponge then wipe dry The Salter Labs 7100 350c c Bubble Humidifier and the B F Medical Bubble Humidifier are suitable for use with the Oxygen Concentrators Or a humidif...

Страница 7: ...sulting in high purity oxygen at flow up to five liters per minute Unlike liquid or cylinder oxygen the concentrator does not require refilling and operates at low pressure The device only requires mi...

Страница 8: ...Indicator Over 85 3 GREEN Indicator light 70 3 to 85 3 YELLOW Indicator light Below 70 3 RED Indicator light Flow Range Maximum Recommended Flow 0 5 LPM 5 LPM Average Power Consumption 320W Pressure R...

Страница 9: ...ut notice 2 This product complies with the following standards C22 2 No 601 1 M90 UL Std No 2601 1 and IEC EN 60601 1 1 3 For electromagnetic compatibility the standards are based as follow CSA 601 1...

Страница 10: ...er additional cartons are available from Merits 7 1 UNPACKING 1 Check for any obvious damage to the carton or its contents If damage is evident notify the carrier or your local dealer 2 Remove all loo...

Страница 11: ...rking objects 8 2 PLUG IN POWER CORD Plug the power cord into a grounded 15 AMP power outlet 8 3 CONNECT HUMIDIFIER if prescribed by your physician Fill the humidifier with distilled water according t...

Страница 12: ...ONNECT OXYGEN TUBING Attach the oxygen tubing to the oxygen outlet of the concentrator or the humidifier bottle if prescribed Attach the prescribed oxygen mask nasal cannula or oxygen catheter to the...

Страница 13: ...he flowline for the desired flowrate 1 18 9 0 1 1 1 2 3 4 5 6 7 5 7 2 FIGURE 5 ADJUSTING THE FLOWRATE 8 7 OXYGEN PURITY INDICATOR This feature monitors the purity of oxygen generated by the oxygen con...

Страница 14: ...gen Concentration Sensor This sensor provides the indication of oxygen purity as follows RED LIGHT Oxygen concentration level falls below 70 3 Discontinue use and call for service obtain alternative s...

Страница 15: ...binet filter located at rear side of the cabinet FIG 6 FIGURE 6 CABINET FILTER LOCATION 1 Remove the cabinet filter and CLEAN AT LEAST ONCE A WEEK depending on environmental conditions FIG 7 2 Clean t...

Страница 16: ...ps long pause Concentrator NOT operating while Power switch ON System Failure 1 Unit overheating due to blocked air intake 2 Internal repairs required 1 Remove and clean cabinet filter 2 Move oxygen c...

Страница 17: ...aid by the purchaser Note There is no other express warranty Implied warranties including those of merchantability and fitness for a particular purpose are limited to the duration of the express limit...