STEP 7

 To inflate the balloon, squeeze the 

trigger allowing the plunger to return to the 
resting position (0 ATM/BAR). Release grip on 
the trigger, locking the plunger into position.

STEP 1

 

Prior to use, free the plunger tip 

by twisting the device plunger/handle 360 
degrees clockwise.

STEP 3

 Remove any excess air by orienting the 

device upwards, squeezing the trigger located 
on the device handle and pushing the handle 
forward. Push forward until the device plunger 
tip is oriented with the black arrow printed on 
the device barrel (approximately 35 mL).

1

2

3

4

7

8

9

5

6

STEP 2

 

To prep device, squeeze the trigger 

and pull back on the handle to aspirate up to 
60 mL of contrast solution into the syringe.

CAUTION

: Inspect the device tubing and stopcock 

(if used) to ensure there is no air in the system.

STEP 4

 Prepare and test the balloon catheter 

according to the catheter manufacturer’s 
instructions for use. 

NOTE

: Refer to the manufacturer’s directions 

accompanying the balloon dilation catheter, or other 

interventional device, for specific information on 

use, maximum inflation pressure, precautions, and 

warnings for that device.

STEP 5 

Create a fluid-fluid connection between 

the balloon and the inflation device extension 
tube and connect the luer connectors securely.

STEP 6

 Squeeze the trigger and pull back on 

the handle to apply vacuum to the balloon.

STEP 8

 To increase pressure, rotate the handle 

clockwise until the desired inflation pressure is 
reached. The locking mechanism maintains the 
pressure.

CAUTION

:  Do not exceed 12 ATM.

NOTE

: Loss of pressure may indicate a leak in the system.

STEP 9

 To deflate the balloon, rotate the 

handle counter-clockwise to relieve pressures 
above 6 ATM. Squeeze the trigger and pull 
the handle back to generate negative pressure. 
Release the grip on the trigger to lock the 
plunger in the negative pressure position.

Prep the Device

Attach the Inflation 

Device to the Balloon

Inflate and Deflate 

the Balloon

E X PA N D I N G   T H E   P O S S I B I L I T I E S ™

endotek.merit.com   1-800-35-MERIT

403150001/A     ID 010115

Instructional Chart

Please refer to the Instructions for Use included 

with the product for complete instructions.