Maintenance
3.1 GENERAL INFORMATION
Meridian Laboratory Ultra High-Speed Brushless Slip Rings contain maintenance-free ball bearings. No
additional lubrication is required.
When in-use on a high-speed system, it is recommended that the slip ring undergo a routine maintenance
check every two to three years. However, if the slip ring is being used for a very high-speed application, more
frequent maintenance may be necessary.
Do not attempt to dismantle the slip ring for any reason.
Please return the ROTOCON UHS Series Slip Ring directly to the manufacturer for proper disposal and
recycling.
If the slip ring has not been in use for a long period of time the dynamic transfer resistance may increase
above the specifications for the unit. This is caused by a clogging of the cells due to deposits of mercury oxide
within the slip ring. The slip ring should be returned to Meridian Laboratory for cleaning and maintenance.