User’s Manual Dwg No 9R141-IR

 

 

STEP BY STEP:  P3, DENSITY OF WATER REFERENCE TEMP 

Step Action 

Display 

Gauge should be in the normal 

Measure Mode

. Normal 

Pressure 

Display. 

Press the 

PRGM/Enter

 key. 

If the lockout is active, the gauge will now prompt for entry of 
the lockout code.  Refer to “STEP BY STEP:  ENTERING A 
LOCKOUT CODE (When Prompted)” on page 22. 

The 

PRGM

 annunciator shows at the top of the 

display, and the display shows the register name 

P0

”. 

Press the 

Up

 or 

Down

 keys repeatedly to scroll 

to the desired register. 

The Zero function (

Up

 

and

 

Down

) will reset the scroll to P0 

and allow selection to continue. 

Display shows  “

P3

”. 

Press 

PRGM/Enter

 to open the register. 

The display shows the current value. 

Press the 

Up

 or 

Down

 keys repeatedly to scroll 

to the desired register value. 

The Zero function (

Up

 

and

 

Down

) will reset the scroll to 

default and allow editing to continue. 

All available choices are scrolled. 
“20”  = 20 Deg. C 
“4” 

= 4   Deg. C 

“60”  = 60 Deg. F 

Press the 

PRGM/Enter

 key to accept the desired 

value. 

The value is accepted, and the register is closed.  
Display shows  “

P3

”. 

Press the 

Backspace

 key to activate the register 

setting and return to 

Measure Mode

Normal Pressure Display. 

Notes 

1.

 

After steps 2 and 3, the 

Backspace

 key will abort the process and return to 

Measure Mode

2.

 

After steps 4 and 5, the 

Backspace

 key will abort the edit and close the register. 

3.

 

After step 6, the 

PRGM/Enter

 key will again open the register for edit. 

4.

 

During any programming operation, if there is no keypad activity for approximately 1 minute, the 
operation is aborted and the gauge returns to 

Measure Mode

 unchanged. 

 

11.

 

P4 – User Scale   All Models 

 

The user scale feature is provided in addition to the standard Engineering Unit choices, to allow the display of 
measured pressure in any Engineering Unit.  The full scale value of the sensor is linearly re-scaled to the value 
entered in this register.  This is particularly useful for displaying linear tank levels in volumetric or mass units. 
 
For example, a 20 PSIG gauge is programmed to read out in feet of water head by entering 46.22 in the P4 
register.  This sets the gauge to read from 0 to 46.22 feet of water, when User Units are selected as the 
Engineering Units    (46.22 feet of water at 20°C is equal to 20 PSIG). 
 
User Engineering Units are selected when ALL Engineering Units are flashing during the Engineering Unit 
selection process.  Once selected, in 

Measure Mode

, all Engineering Units are cleared from the display to 

denote user-units. 

 
 

Содержание M2110P

Страница 1: ...Loop models have outputs that can be used for control or indicating recording functions Both models also have RS 232 communications capability In addition to configuring the gauge the RS 232 port can...

Страница 2: ...truction manual This is the Safety Alert symbol This symbol indicates a WARNING Warnings alert you to actions that can cause personal injury or pose a physical threat Please read these carefully This...

Страница 3: ...Damp Rate All Models 13 10 P3 Density of Water Reference Temperature All Models 14 11 P4 User Scale All Models 15 12 P5 SETPOINT Options Set Point Current Loop Models Only 17 13 P6 P7 SETPOINT Set1 S...

Страница 4: ...Pressure Limits listed in the Product Specifications section of this manual Failure to operate within the specified pressure limit could result in death or serious injury Do not exceed the Maximum Inp...

Страница 5: ...rsion will typically be between 10 to 14 conversions per second depending on the various operating conditions To make the display easier to read only every fourth sample is updated on the display This...

Страница 6: ...re restored and activated 5 Measure Mode displays pressure in the last Engineering Unit used Program Mode Backspace function From Program Mode this key will exit to Measure Mode When editing a registe...

Страница 7: ...ter key accepts and stores the value and closes the register When editing a multi digit value each key press will accept the current digit and proceed to the next until finally accepting the complete...

Страница 8: ...el only P2 Damp Rate Exponential damping time in seconds 0 1 0 2 0 5 1 2 5 10 15 25 50 0 1 No Damping P3 Density of Water Reference Temperature Sets reference temperature for water density when gauge...

Страница 9: ...Press the Up or Down keys repeatedly to scroll to the desired Engineering Unit 3 The Zero function Up and Down will reset the scroll to the first unit and allow selection to continue The current Engi...

Страница 10: ...K The M2110P can be zeroed only when the applied pressure is within 5 of full scale If the applied pressure is greater than 5 of full scale an error code will be displayed when zeroing is attempted ST...

Страница 11: ...ed If the operator action was changing Engineering Units or re zero the gauge the gauge will simply return to normal Measure Mode operation without accepting any change If the operator action was ente...

Страница 12: ...pproximately 1 minute the operation is aborted and the gauge returns to Measure Mode unchanged 8 P1 Time Out Value Battery Models Only This register sets the length of time in minutes for automatic sh...

Страница 13: ...or edit 4 During any programming operation if there is no keypad activity for approximately 1 minute the operation is aborted and the gauge returns to Measure Mode unchanged 9 P2 Damp Rate All Models...

Страница 14: ...ackspace key to activate the register setting and return to Measure Mode Normal Pressure Display Notes 1 After steps 2 and 3 the Backspace key will abort the process and return to Measure Mode 2 After...

Страница 15: ...eturn to Measure Mode Normal Pressure Display Notes 1 After steps 2 and 3 the Backspace key will abort the process and return to Measure Mode 2 After steps 4 and 5 the Backspace key will abort the edi...

Страница 16: ...ll reset the entire register to 0000 and allow editing to continue The decimal point will not be affected and same digit will be flashing _xxx where _ is the digit being edited flashing 6 Press the PR...

Страница 17: ...for edit 6 During any programming operation if there is no keypad activity for approximately 1 minute the operation is aborted and the gauge returns to Measure Mode unchanged 12 P5 SETPOINT Options Se...

Страница 18: ...alue Press the Up or Down keys repeatedly to scroll to the desired register value 5 The Zero function Up and Down will reset the scroll to default and allow editing to continue All available choices a...

Страница 19: ...registers are entered in the current Engineering Unit used by the gauge The current Engineering Unit is illuminated during edit of these registers For example if the gauge is set to read in InH2O at 2...

Страница 20: ...splay shows P6 or P7 Also SET1 or SET2 respectively is displayed to help identify the register in addition to 4 20mA for Current Loop Models 4 Press PRGM Enter to open the register The display will sh...

Страница 21: ...value in its corresponding P6 and P7 registers on increasing pressure The relay will de energize at a value equal to the corresponding P6 and P7 register minus the deadband value on decreasing pressu...

Страница 22: ...a lockout code in register P0 See STEP BY STEP P0 LOCKOUT CODE on page 11 After pressing a key on a locked gauge for a desired function ZERO ENG UNITS PRGM L 00 will appear on the display with the fi...

Страница 23: ...cation for 9600 baud 8 data bits no parity one stop bit and no handshaking The terminal should be able to display at least 24 lines and 70 columns THE MENU All programmable items of the gauge are avai...

Страница 24: ...ION Use of the serial port service is independent of the mode of the gauge and vice versa It is not necessary to change to Program Mode to use any of the menu s features including changing Engineering...

Страница 25: ...gauge The resulting data rate will typically be about ten 10 conversions per second depending on various operating conditions Note that the data sent to the terminal will be damped according to the d...

Страница 26: ...isable the output simply set programmable register P5 to 0 E005 Requested User Units Full Scale range is less than or equal to 0 0 Value must be greater than 0 0 E006 User Entered incorrect Lockout Co...

Страница 27: ...essure connection The SetPoint model utilizes the multi function terminal strip shown at the left This terminal strip has a NEMA 1 rating SPDT relays are not powered by the gauge internally Jumpers fr...

Страница 28: ...ten the 1 8 MNPT pipe thread Do not tighten the fitting without using a wrench on the manifold DN DI Manifold DI Manifold GI CI and AI Manifold Optional Flushing Ports Current Loop Model Wiring 24V 50...

Страница 29: ...e through the front of the panel 5 From the rear insert the panel mount jacks in the grooves on the side of the gauge and slide them firmly against the panel 6 Replace the two 6 32 socket head screws...

Страница 30: ...OWER SetPoint model is standard with selectable 110V AC 50 60 Hz 240V AC 50 60 Hz or 24V DC power Battery models are powered by two Lithium D Cells Battery Life is approximately 100 days at a 100 duty...

Страница 31: ...Scale Only 2 DI units will measure to 20 or 150 PSI whichever is greater pressure relative to absolute zero and 120 of full scale pressure Calibration certification is for 0 F S Full Scale Only 3 GI...

Страница 32: ...nd clearly marked on the return packaging the unit will be returned at the shipper s expense An RMA number will be provided by the Meriam Repair Department when you call fax or e mail your information...

Отзывы: