Operation: 

This unit is not suitable for use above 2000m sea level. Humidity in testing 

environment should be less than 80%RH temperature 0-40°C. 

 

-

 

Switch on the unit 

-

 

Select desired measuring range ideally UNDER and OVER should 
not be display.   

-

 

Select sensitivity weight, for general noise use A, for low 
frequency noise use C. 

-

 

For measuring short burst noise set time weight to fast, for 
normal background noise use slow. 

-

 

For monitor peak noise level, press Max button, only the highest 
on record will be displayed.  

 
Calibration: 

All unit are calibrated from the factory, however we recommend device to 
be recalibrate once a year. To calibrate, insert the mic into the acoustic 

calibrator. Set Sensitivity weight to A, time weight to fast and range to 60-

110dB. After the calibrator is on, use a fine flat screw driver to adjust the 
potentiometer on top of the battery compartment until display 94dB. 

 
Accessories: 

 

 

Instruction manual 

 

Heavy duty protection box 

 

4 x AA battery 

 

1 x Foam windshield 

 

Specification: 

 

Measure Range  

–  

30-130dBA, 35-135dBC 

Accuracy   

–  

+/-1.5dB (under reference condition)  

Frequency Range   – 

31.5Hz – 8.5kHz 

Level Range  

– 

30-80, 50-100, 60-110, 80-130, 30-130dB 

Linearity Range  

– 

50dB/100dB 