background image

 

350.135UK User Manual 

 
The alarm is designed to record maximum readings and average readings over the last 24-

hour period. This allows you to keep track of the presence of carbon monoxide without 
needing to be present in the room. To bring up the reading, simply press and hold TEST 

button until “Ph=” is displayed on the front screen. 

 

It will move to the next page to show the peak reading for the past 24 hours automatically 

after a short wait, or you can manually press the TEST button to enter the next page if you do 
not wish to wait. The following page will display “24H” to indicate it will display the average 

reading next. Press the button again or wait and the average reading will show. The “PPM” 
will flash on the average reading display to differentiate it from the peak reading. 
The following menu option allows you to clear the record. When “clr” appears on the screen, 
simply press and hold the TEST button until “P--” is displayed which indicates the record has 

been deleted. 

 

Maintenance: 

In order to maintain the sensitivity of the alarm, dust needs to be removed from the surface 
on a regular basis. To dust use a slightly damp cloth, a brush or brush attachment of a 
vacuum cleaner. Do not use any solvent based cleaner as the sensor is sensitive to all active 
gas including chemical cleaners. Always perform a manual test after cleaning to ensure the 
alarm is working normally. Also, the battery compartment needs to be checked on regular 
basis for signs of battery acid leakage. Remove batteries immediately if any signs of leakage 
occur. Clean the compartment with a clean tissue/cloth before replacing with new batteries.  
 

Avoid below action to prevent the alarm from damage:

 

 

Do not allow the alarm to come into contact with any coloured or heavy smelling gas as 

this will contaminate the sensor and false readings will occur.  

 

Do not spray aerosols near the alarm, such as: air freshener, insect killer, perfume, hair 

spray etc. as this will result in false readings or contamination of the sensor. 

 

Do not paint over the alarm, always remove alarm from the bracket before painting and 

wait for paint to be fully dry before reinstalling the alarm. 

 

Avoid shock or impact on the alarm, if accidentally dropped check if the battery connection 

is still intact and perform a manual test to ensure the alarm is working correctly. 

 

Do not attempt to repair or modify the alarm. Any modification may cause alarm to 
malfunction and fail to operate in a life-threatening situation, therefore modification and 

repair is strictly restricted.  

In the event of contamination, leave alarm in fresh air for a minimum of 30 mins to up to a 
week to see if it recovers. If readings are still false replace the alarm immediately. 

Содержание COD100B

Страница 1: ...COD100B Carbon Monoxide Alarm Item ref 350 135UK User Manual...

Страница 2: ...ntilation is lacking Please read through the following manual carefully on how to correctly install operate and maintain the device and also for action to be taken in the event of detection of dangero...

Страница 3: ...level It is advised to locate the alarm near the potential source of CO such as rooms with a boiler fireplace or gas oven For optimum detection the alarm needs to be installed at least 1 5m above the...

Страница 4: ...roundings When the alarm is powered the green power LED will flash once every 45 seconds to indicate it is working When the working voltage drops below 3 5V the alarm will beep when the power LED flas...

Страница 5: ...d be below 25ppm Therefore when the count is below 25 the LCD will only display 0ppm and the temperature reading in a 30 second cycle To meet EN50291 standards the alarm will activate within following...

Страница 6: ...o all active gas including chemical cleaners Always perform a manual test after cleaning to ensure the alarm is working normally Also the battery compartment needs to be checked on regular basis for s...

Страница 7: ...an open door or window for fresh air Do a head count to check all persons are accounted for 3 DO NOT re enter the premises or move away from the door window until the alarm has stopped 4 Call the UK E...

Страница 8: ...led appliances in enclosed spaces Do not use fueled generators in enclosed space Paraffin flueless heaters should not be used in a bedroom or an enclosed space Ensure all flues chimneys or other venti...

Отзывы: