600.100UK 

         MTM01 User Manual 

 

Replace the battery as soon as the battery indicator   appears. With a 

low battery, the meter may produce false readings that can lead to 
electric shock and personal injury. 

 

Remove the connection between the testing leads and the circuit being 

tested and turn the meter power off before opening the meter case. 

 

The internal circuit of the meter shall not be altered at will to avoid 

damage of the meter and any accident. 

 

A soft cloth and mild detergent should be used to clean the surface of 

the tester on a regular basis. No abrasive and solvent should be used 

to prevent the surface of the tester from corrosion or damage. 

 

The tester is suitable for indoor use only. 

 

Turn the tester power off when it is not in use and take out the battery 

when not using for a long time. Check the battery regularly; replace the 
battery immediately if any signs of leaking appear. Battery acid will 

damage the tester. 

 

General Specifications 

 

Max display:  

 

LCD (1999 count) 67 x 42mm 

Polarity:    

 

Automatic, indicated minus, assumed plus 

Measure method:    

double integral A/D switch implement 

Sampling speed:    

2 times per second 

Over-load indication:  

“1” is displayed 

 

Operating Environment:  

0ºC-40ºC, at <80%RH  

Storage Environment:  

-10ºC-50ºC, at <85%RH 

Power:   

 

9Vdc (1 x PP3 battery supplied) 

Low battery indication:  

“ ”

 

Dimensions:  

 

190 x 90 x 33mm   

Weight:    

 

190g (including battery) 

 
 

 

 
 

 

Содержание 600.100UK

Страница 1: ...Item ref 600 100UK MTM01 Digital Multimeter User Manual...

Страница 2: ...age as marked on the tester between the terminals or between any terminal and grounding The rotary switch should be in the right position and no changeover of range shall be made while measurement is...

Страница 3: ...ld be used to prevent the surface of the tester from corrosion or damage The tester is suitable for indoor use only Turn the tester power off when it is not in use and take out the battery when not us...

Страница 4: ...600 100UK MTM01 User Manual LCD Power Button Range switch A mA terminal V terminal 10A terminal COM terminal HOLD Button Multi functionadapter...

Страница 5: ...200mV 100uV 0 5 of rdg 3D 2V 1mV 20V 10mV 0 8 of rdg 5D 200V 100mV 600V 1V 1 0 of rdg 5D Input Impedance 10M Overload Protection 1000V DC or 750V AC rms Max Input voltage 1000V DC AC Voltage RANGE RES...

Страница 6: ...urrent RANGE RESOLUTION ACCURACY 2mA 1uA 1 8 of rdg 2D 20mA 10uA 200mA 100uA 2 0 of rdg 2D 10A 10mA 2 0 of rdg 10D Overload Protection mA F0 5A 600V fuse 10A F10A 600V fuse Voltage Drop 200mV AC Curre...

Страница 7: ...4D 20k 10 200k 100 2M 1k 20M 10k 1 0 of rdg 10D 200M 100k 5 rdg 10 10D Open Circuit Voltage about 3V Overload Protection 250V DC AC rms Capacitance RANGE RESOLUTION ACCURACY 2nF 1pF 4 0 of rdg 5D 20nF...

Страница 8: ...ar on Digital Display along with the voltage polarity Please note In small range the meter may display an unstable reading when the test leads have not been connected to the load to be measured It is...

Страница 9: ...e When the display shows the over range symbol 1 a higher range must be selected In addition 10A function is designed for intermittent use only RESISTANCE MEASUREMENT 1 Connect red lead to V black lea...

Страница 10: ...ance The tested capacitor should be discharged before the testing procedure Never apply voltage to the input or serious damage may result CONTINUITY TEST 1 Connect the BLACK test lead to the COM jack...

Страница 11: ...of the adaptor 4 LCD display will show the approximate hFE value BATTERY AND FUSE REPLACEMENT 1 Battery and fuse replacement should only be done after the test leads have been disconnected and power...

Страница 12: ...nd black 9V PP3 battery EN61010 1 2010 This product is classed as Electrical or Electronic equipment and should not be disposed with other household or commercial waste at the end of its useful life T...

Отзывы: