Insta-Therm Service Manual 

TROUBLESHOOTING 

 

SYMPTOM POSSIBLE 

CAUSE 

EVALUATION 

 

Incoming Power Supply 

 

Check breakers, reset if required. Check power plug 
to be sure it’s firmly in receptacle. 
Measure incoming power 208v 

 

 

Hi Limit Thermostat 

 

Terminals are normally closed, if open push in reset 
button. 
If button will not reset replace hi limit. 
Hi limit opens at 60C 

 

 

Unit will not turn on 
 
 

 

On – Off Switch

 

 

Check continuity between switch terminals. 
Replace switch as needed. 

 

 

Contactor 

 

Check for supply voltage to contactor contacts, if no 
voltage present trace back to on-off switch. 
If voltage is present, check to insure contacts are 
closing, replace as needed. 

 

 

Solid State Relay  
(Current limiting) 

 

Check for supply voltage to relay contact, if no voltage 
present trace back to contactor. 
Check for output voltage, if no output voltage check 
percentage dial on relay is set to 50%. 
If no output voltage present replace as needed. 

 

 

No Stand By Heat 

 

Elements 

 

Check the amp draw on each element for proper load. 
Check specification plate for rating information. 
Check individual elements for opens, shorts and 
proper resistance 56ohms. 

 

 

Micro Switch 

 

Check continuity of the NC held open switch with 
sufficient weight on surface, if no continuity, replace 
switch. 

 

 

Solid State Relay  
(Current limiting) 

 

Check for supply voltage to relay coil (this will be 
indicated by two lamps on the unit) if no voltage 
present trace back to contactor and/or the micro 
switch. 
Check for output voltage, if no output voltage check 
percentage dial on relay is set to 50%. 
If voltage is present at coil check contacts are closing, 
replace as needed. 

 

 

No Full Power 

 

Elements 

 

Check the amp draw on each power leg for proper 
load. Check specification plate for rating information. 
Check individual elements for opens, shorts and 
proper resistance 56ohms. 

 

 

Hi Limit Trips 

 

Cooling Fan Motor 

 

Check to see cooling fan is operational, if not check 
for voltage to motor, if no voltage present trace back 
to on – off switch, if voltage is present replace motor. 

 

 

Содержание Insta-Therm 16000 Series

Страница 1: ...0136 REV 8 2 10 Merco LLC 1111 North Hadley Road Fort Wayne Indiana 46804 Telephone 260 459 8200 Fax 888 790 8193 Technical Support 800 678 9511 mercosavory com PARTS SERVICE MANUAL Insta Therm 16000...

Страница 2: ...s Amps Watts Hz Phase Cord Plug Weight 16001 208 11 25 2340 60 Single 6 NEMA 6 20P 47 lbs 21 3 kg 16003 230 9 91 2280 50 Single CEE 7 7 47 lbs 21 3 kg 16005 208 11 25 2340 60 Single 6 NEMA 6 20P 47 lb...

Страница 3: ...Insta Therm Service Manual 3 SCHEMATIC MODEL 16001...

Страница 4: ...Insta Therm Service Manual 4 SCHEMATIC MODEL 16003 16005...

Страница 5: ...through the high limit thermostat Also to the open contacts of the main power relay HEATING CIRCUIT Closing the main power switch feeds power to the cooling fan and the coil of the main power relay I...

Страница 6: ...replace as needed No Stand By Heat Elements Check the amp draw on each element for proper load Check specification plate for rating information Check individual elements for opens shorts and proper r...

Страница 7: ...Insta Therm Service Manual 7 This page intentionally left blank...

Страница 8: ...D 340180 Cordset 240V 20A w sleeve sp E 340032 Bushing Heyco F 340117 Hi Limit G 340033 Contact Coil H 369331 Finger Guard I 369378 Fan Motor J 340036 Solid State Relay K 340037 Terminal Block X3 L 2...

Страница 9: ...Insta Therm Service Manual 9 EXPLODED VIEW ILLUSTRATION...

Страница 10: ...support bracket in place and remove the bracket Retain the screws for later reinstallation Leave the right hand side heater support bracket in place CAUTION Heater elements must not be handled with ba...

Страница 11: ...ght hand wall of the unit feeding the lead wires through the hole in the left hand wall and heater support bracket The small diameter of the ceramic end caps on the element should fit snuggly into the...

Страница 12: ...the heater element feeding the lead wires through the holes Secure using the screws previously removed STEP 5 CONNECT THE LEFT SIDE HEATER WIRES Connect all four 4 heater wires on the left side to the...

Страница 13: ...to the terminal block as shown Heater wire 1 and 2 are connected to terminal block position 3 heater wire 3 to terminal block position 2 and heater wire 4 to terminal block position 1 STEP 7 REINSTAL...

Страница 14: ...Insta Therm Service Manual 14 This page intentionally left blank...

Страница 15: ...Insta Therm Service Manual 15 This page intentionally left blank...

Страница 16: ...Insta Therm Service Manual 16...

Отзывы: