background image

TOC INDICATOR OPERATION

The Milli-Q® Reference A+ system uses an integrated 

185/254 nm UV lamp to oxidize neutral organics into 

charged molecules and facilitate their removal by the 

mixed bed ion-exchange resin in the polishing cartridge. 

The TOC Indicator employs this same UV lamp to indicate 

TOC value by measuring the difference in ultrapure water 

resistivity upstream and downstream from the UV lamp.

Previous attempts to use this method resulted in devices 

with an important drawback: they did not take into account 

the fact that all organic molecules are not oxidized at the 

same speed and that the nature of organics in feed water 

may change depending on the geographical location and 

the time of year. These TOC indicators were calibrated in 

factories using feed water containing a specific mix of 

organic substances—and they performed reasonably well—

with the same feed water. However, they were unable to 

adapt their operation to feed water variations, and often 

delivered erratic results when the mix of neutral organic 

molecules in the feed water was changed. Consequently, 

these devices could very well indicate low product water 

TOC in cases where the TOC value was actually high.

On the other hand, the 

TOC Indicator

 in the 

Milli-Q® 

Reference A+ system

 uses an operating process that 

includes a patented “TOC Curve Check” phase, which 

adjusts the TOC analysis results each night to the evolution 

of the mix of organics in the system’s feed water. The same 

patented process allows taking into account possible 

variations of UV light intensity that might be due to 

deposits on the quartz sleeve surrounding the lamp.

Merck Millipore’s patented process ensures that the TOC 

values displayed by the Milli-Q® Reference A+ system are 

reasonably accurate for TOC levels within 1 and 10 ppb. 

The system will also indicate any increase in TOC values 

above this level.

Millipore, Milli-Q, A10, Q-Gard and Quantum are registered trademarks of Merck KGaA Darmstadt Germany. The M Mark and Merck 
Millipore are trademarks of Merck KGaA. 
Lit. No. PF2954EN00. 
© 2012 EMD Millipore, Billerica, MA. U.S.A. All rights reserved.

TOC INDICATOR AND MONITOR

The differences between a Merck Millipore TOC Indicator 

and a Merck Millipore TOC Monitor (A10) are listed in the 

table below.

Parameter

TOC 

Indicator TOC Monitor

Monitoring

Frequency

Once per

second

Once every 5 minutes

- N.B. This is not very

important as TOC 

values typically do not 

change very quickly

Accurate TOC

Detection Range

1 - 10 ppb 1 - 999 ppb

TOC Values Display 1 - 

999 ppb

1 - 999 ppb

Calibration to

reference solutions NO

YES

Certificate

of Calibration

NO

YES

TOC Curve Check

YES

NO - The TOC curve 

check process is not 

required as the TOC 

monitor is calibrated 

to reference solutions

Suitability test

as required by USP NO

YES

Cost

Low

Medium

2

TOC INDICATOR

During normal system operation, water at a 

resistivity of 18.2 MΩ·cm @ 25 °C flows through 

the 185/254 nm UV lamp. Water conductivity and 

temperature are measured at the outlet of the UV 

lamp whenever water is recirculated or delivered. 

Oxidation by the UV lamp of the neutral organics 

into charged molecules generates a conductivity 

increase whose value is converted to a TOC value 

by an algorithm. To ensure that the algorithm takes 

into account any variations in the nature of the 

water’s organic compounds or possible decrease in 

UV light intensity, a “TOC Curve Check” is performed 

every day.

TOC CURVE CHECK

Once a day, or when the user requires, the system 

performs a “TOC Curve Check.” During this operation, 

the water flow is diverted from the UV lamp. This 

allows the system (1) to check that the water resistivity 

entering the lamp is still at 18.2 MΩ·cm @ 25 °C and 

(2) to capture a water sample inside the UV lamp. 

The first sample will be oxidized for 20 seconds, then 

released and its conductivity measured. This process is 

repeated several times with increasing oxidation times 

(the second sample is oxidized for 30 seconds before 

release and measurement, the third for 40 seconds, 

etc.). The data collected are used to build a curve of 

the conductivity as a function of oxidation time. 

During the day, this curve is used to generate valid 

TOC measurements based on the small conductivity 

variation resulting from the passage of the water 

through the UV lamp.

Отзывы: