78
Operating Instructions - CPC4000
Industrial Pressure Controller
CPC4000
7.9.2 CPC4000 Error Codes
Code
Serial Poll Byte Description
Error String Returned
E00
00h
No errors
NO ERRORS
E05
45h
Parameter error
EGPIB PARAMETER ERROR: String
that was sent
E07
47h
Syntax error
EGPIB SYNTAX ERROR: String that
was sent
7.9.3 SCPI Commands and Queries
The WIKA SCPI command is compatible with older WIKA instruments such as the original CPC8000
Notes:
1. Note that WIKA and Ruska 7010 Emulation return slightly different responses to some queries.
2. Except for the engineering units selection, the numeric suffix selects the applicable transducer [R]:
1 = Sensor 1
2 = Sensor 2
This numeric suffix [R] always defaults to the active transducer if not specified.
Command
Response/Function
STATus
:OPERation
:CONDition?
Returns an integer value representing instrument
status that can be decoded.
Bit 0: Zeroing active
Bit 1: Control Setpoint has not been reached.
Bit 2: Reserved 0.
Bit 3: Reserved 0.
Bit 5: Measuring. The instrument is actively mea-
suring
MEASure
[:PRESsure][R]?
Returns the pressure from range R
:TEMPerature[R]?
Returns the temperature from range R
:RATE[R]?
Returns the rate/sec from range R
:BAROmetric?
Returns the barometric pressure
CALibration
[:PRESsure][R]
:MODE?
Returns 1=calibrated or 0=not calibrated
:DATE?
Returns date of cal “MM/DD/YY”
:DATE <i,i,i>
Sets date of cal YYYY,MM,DD
:ZERO?
Returns zero offset
:ZERO <n>
Sets the zero offset