Memmert WNB 10 Скачать руководство пользователя страница 16

BASIC

page 16

12 Check list for fault rectifi cation

In case of a malfunction contact an authorized service station for Memmert equipment or please inform 

the Memmert service department (see section 16).
In case of queries always specify model and serial number (on the rating label).

13 Action on supply failure

supply failure

After a failure of the supply, operation continues with the previously set parameters.

14 Glossary

Glossary

• 

nominal temperature = themaximum adjustable setpoint temperature of the bath.

• 

ambient temperature = the continuous temperature of the room in which the bath is set up.

Main switch ON,

no indication on the display

Main fuse 15A or instrument fuse T80mA 250V~ on circuit board 

55167.x has blown

Controller faulty

Electrical supply interrupted

 symbol not alight

Ambient temperature too high

Temperature in bath above the selected setpoint temperaturer

 symbol alight

Temperature protection (TB) has operated

Liquid level too low

 symbol fl ashes

Monitoring relay has operated

C O N F

Error on self test

E - 1

Power module triac faulty

E - 2

Power module faulty

E - 3

Pt100-temperature probe faulty

E - L

Communication to power unit interrupted

Содержание WNB 10

Страница 1: ...Waterbath WNB 7 45 OPERATING INSTRUCTIONS...

Страница 2: ...BASIC page 2...

Страница 3: ...s under warranty If these operation and for any possible claims under warranty If these Instructions are disregarded all claims under warranty guarantee Instructions are disregarded all claims under w...

Страница 4: ...296 Housing height with gable cover G mm 337 337 347 405 343 401 Weight kg 11 14 16 17 24 26 Current consume A 5 2 5 2 7 8 8 7 10 4 12 2 Power W 1200 1200 1800 2000 2400 2800 Ambient conditions Ambien...

Страница 5: ...tic compatibility electromagnetic compatibility This product is intended to operate on a supply network with a system impedance Zmax at the transfer point building connection of 0 292 Ohm max The user...

Страница 6: ...ately as accessory at cover with concentric ring sets gable cover to drain condensate cooling device various racks for test tubes bottles etc 3 Installation The unit must be placed on a horizontal non...

Страница 7: ...ever use demineralised water with a conductance level of 5 10 microsiemens and a pH value between 5 and 7 The tray may suffer damage if ultrapure water or DI water with an electrical conductance level...

Страница 8: ...thus securing an optimal uniform temperature distribution 4 1 Controls and indications relatively largest heat loss since fur thest away from the heated ribs minimum heat loss since very close to the...

Страница 9: ...hes brightly and can now be altered with the push turn control while holding down the SET key protection against unintentional alteration If the push turn control is rotated quickly the setpoint is al...

Страница 10: ...isplay shows CCC The setting boiling mode is not stored permanently After the bath is switched off and on again the controller shows again the previously selected setpoint C C 4 3 2 Delayed switch on...

Страница 11: ...the actual temperature and the running time of the hold time As with switch on delay the residual time is shown with a negative sign and runs down If no hold time is required it can be de activated by...

Страница 12: ...ath heats up to the selected setpoint temperature Heating is indicated by the symbol C Bath holds setpoint temperature During the hold time the symbol ashes and the display alternates between the resi...

Страница 13: ...certain level As a warning the symbol lights up lights up continuously Fault recti cation after the TB cut out has been activated 1 Switch off the bath and allow it to cool down 2 Rectify the fault e...

Страница 14: ...p to the temperature of the thermo uid 9 Level control special equipment Level control special equipment 2 different lling levels can be maintained constant if the unit is tted with a level control sy...

Страница 15: ...ents appropriate for Please use only detergents and antiliming agents appropriate for stainless steel for the cleaning of the tank and the housing stainless steel for the cleaning of the tank and the...

Страница 16: ...oint temperature of the bath ambient temperature the continuous temperature of the room in which the bath is set up Main switch ON no indication on the display Main fuse 15A or instrument fuse T80mA 2...

Страница 17: ...ective DIN EN 61326 2004 05 EN 61326 1997 EN 61326 A1 1998 EN 61326 A2 2001 EN 61326 A2 2003 The designated product is in conformity with the European Low Voltage Directive 2006 95 EEC including amend...

Страница 18: ...of domestic refuse For disposal please contact your dealer or the manufacturer Products which are infected infectious or contaminated with health endangering substances are excluded from return Please...

Страница 19: ...16 H heating 8 hinge bolts 15 hold time of the setpoint temperature 11 I indications 8 initial start up 7 installation 6 L level control system 7 14 liquid level 7 low level protection 13 M maintenanc...

Страница 20: ...Memmert GmbH Co KG Postfach 1720 D 91107 Schwabach Tel 49 0 9122 925 0 Fax 49 0 9122 145 85 E Mail sales memmert com www memmert com 15 06 2018 Waterbath BASIC englisch D10329...

Отзывы: