MEMMERT GmbH + Co. KG
Willi-Memmert-Straße 90-96
D-91186 Büchenbach
Germany
Phone: +49 (0)9122 925-0
Fax:
+49 (0)9122 14585
E-mail: [email protected]
Internet: www.memmert.com
Customer service:
Hotline:
+49 (0)9171 9792 911
Service fax: +49 (0)9171 9792 979
E-mail: [email protected]
© 2017 Memmert GmbH + Co. KG
3
4.
Preparing the wall cut-out
When making the wall cut-out, ensure that its size is between the minimum and maximum
dimensions. If the wall cut-out is too small, the appliance cannot be positioned.
If the cut-out is too large, the cleanroom will not be sealed properly.
The wall has to be at least 100 mm thick and must not be thicker than 300 mm. This is
important to ensure that the service cover on the appliance top is accessible.
If several units are installed next to each other in the wall, the distance between the units
must be at least 300 mm due to the frame width.
The following minimum and maximum dimensions apply for the pass-through ovens
UF160TS, UF260TS, UF450TS and UF750TS, which are usually installed on the floor.
Legend:
Wall
Tolerance zone
Cut-out
Wall cut-out dimensions for an UF160TS appliance:
825 mm
870 mm
22.5 mm
22.5 mm
25 mm
mm
mm
The width allowance is 22.5 mm on each side, i.e. 45 mm in total.
The height allowance is 25 mm.