background image

D39322 | Date 03/2018  

9

Safety regulations

After inhaling CO

After inhaling CO

22

:

High concentrations can cause suffocation. Symptoms may include a loss of mobility and 

unconsciousness. The victim is not aware of suffocating.
Low concentrations of CO

2

 can cause accelerated breathing and headaches.

Anyone affected should breathe fresh air, using a breathing device independent of recircu-

lating air. Keep the person warm and calm. Call a doctor. In case of respiratory arrest, use 

artificial respiration.

If gas is escaping:

If gas is escaping:

Leave the room immediately, warn others and ventilate the room. If you re-enter the room, 

use a breathing device independent of recirculating air if it has not been established that the 

atmosphere is harmless.

1.9   Switching off the appliance in an emergency

Press the main switch on the ControlCOCKPIT

(Fig. 1) and disconnect the power plug. This 

disconnects the appliance from the power 

supply at all poles. 

Hauptschalter

>

Zu der Typenbezeichnung gibt es momentan drei Entwurfsrichtungen, 

ich kann Ihnen noch nicht sagen, ob Memmert hierzu schon eine Entscheidung 

getroffen hat. 

Die hier gezeigt Variante, ist die von uns empfohlene Richtung.

UF 110

PLUS

ON

344

.4

444

4

.

TEMP

Set

°C

°C

FLAP

40

%

TIMER

h

d

10

02

End

Sept.

29

22 24

FAN

%

0

%

100

ALARM

max

444

.4°C

auto

+/-

 

0.0

K

min

444

.4°C

O

O

ON

N

N

N

O

O

N

O

Manu

44.Sept

Fig. 1 

Switch off the appliance by pressing the 

main switch

Содержание ICO105MED

Страница 1: ... memmert com www atmosafe net 100 ATMOSAFE MADE IN GERMANY ICOmed with IVF module ADDITION TO OPERATING MANUAL OPERATING MANUAL OPERATING MANUAL CO2 2 2 INCUBATOR ICOMED WITH IVF MODULE ICO50MED ICO105MED ...

Страница 2: ... service memmert com When calling for service enquiries please state the appliance number given on the nameplate see page 10 Shipping address for repairs Shipping address for repairs Memmert GmbH Co KG Customer service Willi Memmert Str 90 96 91186 Büchenbach Germany Please contact our customer service department before sending appliances for repair or before returning equipment or the shipment ma...

Страница 3: ...configurations and sizes illustrations in this manual may be slightly different to the actual appearance Function and operation are identical Other documents to be observed the operating manual for the ICOmed CO2 incubator For operation of the appliance with MEMMERT AtmoCONTROL observe the separate software manual For service and repair work observe the separate service manual Storage and resale T...

Страница 4: ...module 3 1 Additional package for self installation of the IVF module 3 1 Additional package for self installation of the IVF module 12 3 2 Assembly 3 2 Assembly 3 2 Assembly 12 3 2 1 Cleaning 12 3 2 2 Checking the temperature sensor 3 2 2 Checking the temperature sensor 3 2 2 Checking the temperature sensor 13 3 2 3 Fitting the installation frame 3 2 3 Fitting the installation frame 3 2 3 Fitting...

Страница 5: ...ns are explained below 1 1 1 Terms used WARNING Warns of a dangerous situation that might lead to death or serious injuries CAUTION Warns of a dangerous situation that might lead to moderate or minor injuries NOTICE Warns of material damage 1 1 2 Signs used Danger of elec trocution Fire hazard Gases vapours Danger of frost bite cold burns Warning of gas bottles Danger of burns Disconnect the mains...

Страница 6: ...not emit any poisonous or explosive vapours when heated up WARNING Leaving the door open during operation can cause the appliance to overheat or pose a fire hazard Do not leave the door open during operation WARNING Depending on operation the surfaces in the interior of the appliance and the chamber load may still be very hot after the appliance is switched off Touching these surfaces can cause bur...

Страница 7: ...erman Hazardous Substances Ordinance GefStoffV You should nevertheless familiarise yourself with the applicable safety regulations prior to handling such gas bottles 1 3 Intended use The CO2 incubator ICOmed with IVF module is used to generate and maintain constant ambi ent conditions for the in vitro fertilisation IVF application field especially for the incubation of oocytes spermatozoa and zygo...

Страница 8: ...s and alterations No independent conversions or alterations may be made to the incubator No parts may be added or inserted which have not been approved by the manufacturer Unauthorised changes or alterations result in the EC declaration of conformity losing its valid ity and the incubator may no longer be operated The manufacturer is not liable for any damage danger or injuries that result from un...

Страница 9: ...he room immediately warn others and ventilate the room If you re enter the room use a breathing device independent of recirculating air if it has not been established that the atmosphere is harmless 1 9 Switching off the appliance in an emergency Press the main switch on the ControlCOCKPIT Fig 1 and disconnect the power plug This disconnects the appliance from the power supply at all poles Hauptsc...

Страница 10: ...nd LCD displays for description and opera 2 Main switch tion see ICOmed operating instructions Main switch tion see ICOmed operating instructions 3 Outlet 4 IVF module 5 Markable drawers 6 Unlocking handle 7 Inner glass door Unlocking handle Inner glass door Unlocking handle 2 2 Description The ICOmed CO2 incubator with IVF module is used to generate and maintain constant ambient conditions for th...

Страница 11: ...ed the temperature humidity and CO2 control is interrupted until it closes again 2 3 Standard package Standard package IVF module Drawer insert for Petri dishes ICOmed 50 12 units ICOmed 105 16 units 2 holders for IVF tubes Magnetic foil Custom calibration certificate ...

Страница 12: ...ad screws M4x8 12 units 3 2 Assembly Installation is only necessary if the IVF module is not already installed in the ICOmed incubator at the factory 3 2 1 Cleaning The components for the installation frame are cleaned at the factory by Memmert However shipping and packaging may cause contamination If necessary clean the interior of the cabinet and the individual parts of the installation frame in...

Страница 13: ... sensor is located correctly and if neces sary adjust in position in the holders Fig 3 3 2 3 Fitting the installation frame CAUTION Risk of injury Wear gloves and work shoes Required tools Required tools Allen key sizes 2 5 and 5 1 Position the lower mount ing frame centrally on the bottom of the unit so that the shelf on the front side protrudes approx 10 mm into the interior of the unit The fixi...

Страница 14: ...me on the left and right side over the heating bead ing The side installation frames rest on the lower installation frame 3 Slightly tighten the side mounting frames on both sides with six M4 x 8 pan head screws Ensure an even gap between the side and lower installation frames ...

Страница 15: ...en the side and top bottom mounting frames is approximately equal 7 Test whether the mounting frame is firmly seated o A 3 2 4 Insert IVF module WARNING Risk of injury due to the weight of the IVF module The IVF module 50 weighs approx 25 kg the IVF module 105 approx 35 kg Always hold the IVF module with both hands 1 Grip the IVF module by the frame with both hands and insert it into the tube Fig ...

Страница 16: ... marked with an arrow and pull out the drawer Fig 6 2 The drawer is secured against being pulled out by the mounting bracket mounted in the rear area of the drawer To release the safety device push the lever upwards to release it Fig 7 Fig 6 To pull out a drawer lift the tab Fig 7 Press the safety lever upwards to remove the drawer Fig 7 remove the drawer Fig 7 Press the safety lever upwards to re...

Страница 17: ...ews for fastening in the interior with an Allen key size 5 see Fig 5 2 Remove the upper left or right drawer see section 3 2 5 3 Reach into the drawer opening and pull out the slide in unit until the frame is within reach Fig 8 4 Grip the IVF module by the frame with both hands and pull it out of the tube Fig 9 Fig 8 Pull the slide in unit out of the drawer opening Fig 9 Pull the IVF module out of...

Страница 18: ... very hot inside during the sterilisation process Do no open door If you open the door the on going sterilisation process will be terminated and destroyed NOTICE The sterilisation programme is not intended to sterilise the chamber load but only to sterilise the interior of the appliance The ICOmed IVF incubator is not a steriliser with respect to the Law on Medical Products 4 3 1 Preparation 1 Ope...

Страница 19: ...rawers and the door 6 Attach magnetic foil to the door Fig 12 7 Carry out the sterilisation program as de scribed in the ICOmed incubator s operating instructions NOTICE Do not leave the appliance unattended during the sterilisation process Fig 12 Magnetic foil ...

Страница 20: ...s 5 2 Start the incubation process 1 Wipe off any descriptions on the magnetic foil and the drawers 2 Insert the Petri dishes with the incubation charge 3 Close the door 4 Label the outer magnetic film and drawers with whiteboard pens to assign the samples do not use permanent markers NOTICE Only open the outer door for checking purposes and otherwise keep the door closed When the door is opened t...

Страница 21: ...ork CAUTION Danger of cuts due to sharp edges Always wear gloves when work ing inside the chamber Metal surfaces can be cleaned with normal stainless steel cleaning agents To clean the cham ber interior the interior fittings have to be removed instructions from page 17 Wipe the IVF module with fuzz free cloths and cleaning disinfection agent ...

Страница 22: ......

Страница 23: ......

Страница 24: ...mert GmbH Co KG Willi Memmert Straße 90 96 D 91186 Büchenbach Tel 49 9122 925 0 Fax 49 9122 14585 E Mail sales memmert com facebook com memmert family Die Experten Plattform www atmosafe net ICOmed with IVF module D39322 Date 03 2018 Englisch ...

Отзывы: