Calibration Software PTC04 

User Manual 

Page 12 of 17

    

 

 

REVISION 1.44.01 - 2 MAY 2018 

UI_0PTC04AAMLX 

Check Calibration Setup 

Before the measurements for the calibration check are started, the software will perform a hardware check on the on 
the calibration setup. This test will verify: if the PTC04 is well connected and in good health, if the PTC04-DB-
Calibration is plugged in on the main board, the resistors on the PTC04-DB-Calibration, if the PTC04-DB-Calibration is 
connected to the reference voltmeter and if the reference voltmeter is connected to the PC (this is not the case for 
the manual calibration). If the check fails the calibration check will be discontinued. The following message is 
displayed if a failure is detected on the calibration setup. 
 

 

Figure 19: Check calibration setup failed. 

 

Measurement failed 

The following error message is shown during the calibration procedure;  

 

If there is a bad connection in the setup. For example: between the PTC04 main board and the 
calibration board. 

 

When using a Keithley 2000 multimeter for the calibration, the failure can also be caused by a wrong 
measurement of the Keithley. For example: communication error.  

 

When you are using the manual calibration procedure. The software will perform a check on the data 
you enter. If there is a large deviation from the expected data. The check will fail and the calibration 
check is ended. 

 

 

 
 
 
 
 
 
 

Figure 20: Check Measuring line 1 Failed 

Figure 21: The calibration check has been cancelled. 

 
To determine this: 

 

Press OK on both messages to cancel the calibration check. 

 

Remove the calibration board, place it back and check the connections with the voltmeter. The connection of 
the calibration board and the voltmeter are very imported. A bad connection or any additional load will 
corrupt the calibration parameters.  

 

After checking the setup, restart the calibration check. If the same channel fails again it is possible that there is 
hardware damaged on the PTC04. Please contact your local sales representative in this case. 

 

 

Содержание PTC-04

Страница 1: ...AAMLX User Manual Calibration Software PTC04 Programmer Introduction In this document you ll find a description of The Calibration Setup How to proceed with the calibration check What if the calibrati...

Страница 2: ...r 4 Calibration check 5 Before starting the calibration check 5 Manual voltmeter 8 The calibration check 9 There are three possible outcomes 9 What if the Calibration check failed 11 Check Calibration...

Страница 3: ...sion of the PTC04 has no impact on the quality of the PTC04 calibration it is advised to upload the ptc04_mlx90251_firmware hex or ptc04_mlx90316_firmware hex firmware to the PTC04 before replacing th...

Страница 4: ...VEN 7BITS OR PARITY NONE 8BITS TX TERM CR LANGUAGE SCPI LANGUAGE SCPI For configuration instructions please consult the user s manual of your multimeter All other settings are set by the software via...

Страница 5: ...and compare its results with the calibration parameters If the calibration parameters are out of tolerance they can be corrected Before starting the calibration check It is advised to leave the progra...

Страница 6: ...4 01 2 MAY 2018 UI_0PTC04AAMLX Figure 6 Calibration Check Press Verify Calibration to start the calibration You are reminded to leave your PTC04 powered up for minimum 30 to 60 minutes before starting...

Страница 7: ...With a manual voltmeter the program will start with calibrating the internal voltmeter which will be used to check the PTC04 programmer Figure 8 Proceed Check Press yes to continue At this stage the...

Страница 8: ...internal reference meter Press OK to continue For the calibration of the internal reference meter two series of 32 values are required Once the internal reference meter is calibrated the software wil...

Страница 9: ...At the end of the calibration check the program verifies if a correction of the calibration parameters is needed Figure 14 Calibration Check There are three possible outcomes 1 The program notifies yo...

Страница 10: ...04 If you press no there are no modifications done 3 At the end of the calibration the software verifies if the programmer was at temperature when the calibration started If the programmer was not at...

Страница 11: ...input of measured data If this occurs check the setup and redo the calibration check Very important If the calibration check is done with a manual voltmeter you have to take extra care that the measur...

Страница 12: ...a bad connection in the setup For example between the PTC04 main board and the calibration board When using a Keithley 2000 multimeter for the calibration the failure can also be caused by a wrong mea...

Страница 13: ...calibration parameters In the main window you get a list of the channels that have a large deviation from the default values Figure 23 List of channels that have a large deviation The large deviation...

Страница 14: ...s PTC04UI Calibration details How does the calibration work ADC In the graph below you have two curves the measurements of the ADC left Y axis and the reference measurements of the external voltmeter...

Страница 15: ...TC04 and the External Voltmeter which results in to correction parameters for offset and gain Graph 2 At the end of the calibration check these values of the ADC channels and the DAC are compared with...

Страница 16: ...d then the additional channels that are connected to that source Voltmeter on the PPS1 channel 0 and Channel 16 High sensitivity of the ADC PPS1 channel 0 of the DAC Current meter on the PPS1 channel...

Страница 17: ...Melexis and the use thereof or anything described by the Information does not grant explicitly or implicitly to any party any patent rights licenses or any other intellectual property rights This doc...

Отзывы: