background image

11 Maintenance

64

11 Maintenance

Servicing intervals

Interval

Measure

Device components

Daily

Check for soiling, deposits or
damage

Chamber inc. door seal and chamber
sealing face, door lock, mount for the load

Every 2 months

Clean, check and oil the locking
spindle and nut

Door mechanism

After 24 month
or 4000 cycles

Maintenance

By the authorized customer services
working in accordance with the
maintenance instructions.

As required

Cleaning the surfaces

Housing parts

Cleaning

NOTICE

Inappropriately performed cleaning can lead to the scratching of and damage to surfaces
and the development of leaks in sealing surfaces.

This also favours the development of soiling deposits and corrosion in the sterilization
chamber.

n

Comply with all information regarding cleaning of the part affected.

n

Do not use any hard objects for cleaning such as a metal saucepan cleaner or a steel brush.

Sterilization chamber, chamber sealing face, mount, trays

To maintain the value of your device and to avoid persistent contamination and deposits, MELAG recom-
mends weekly cleaning of the surfaces (e.g. with the MELAG chamber cleaning set).

Cleaning the sterilization chamber, chamber sealing face, mount and trays

The following must be fulfilled or present:

ü

The device has been switched off and the power plug has been unplugged from the socket.

ü

The device has been completely cooled.

ü

Trays or sterilization containers and the associated mount have been removed from the sterilization
chamber.

1.

Wet the surfaces to be cleaned completely with the cleaning fluid. 

PLEASE NOTE

: You should not allow cleaning fluid to enter the piping coming from the sterilization

chamber.

2.

Spread the cleaning fluid evenly with a non-fuzzing cloth.

3.

Allow the cleaning fluid to act and evaporate for a sufficient time.

4.

Using a new non-fuzzing cloth, spread ample amounts of demineralised water over the cleaned
surfaces.

5.

Wipe the surfaces thoroughly to remove cleaning residues. Repeat this process as necessary after
wringing out the cloth.

Ê

Residues of cleaning fluids can ignite or cause deposits on the instruments.

6.

Allow the cleaned areas to dry completely. This may take a few minutes.

7.

Finally, wipe the cleaned surfaces with a dry, non-fuzzing microfibre cloth.

8.

Clean the door seal with neutral liquid detergent.

Содержание Vacuklav 40 B+ Evolution

Страница 1: ...al reliability has established MELAG as the world s leading manufacturer in the instrument reprocessing and hygiene field You our customer are justified in your demand for the best products quality an...

Страница 2: ......

Страница 3: ...vice 14 Load mounts 15 Energy saving key 16 Colour touch display 16 LED status bar 18 5 First steps 19 Setup and installation 19 Feed water supply 19 Supply with cooling water 19 Switching on the stea...

Страница 4: ...Screensaver 60 Log printer MELAprint 42 44 61 Label printer MELAprint 60 61 Sensitivity 62 Energy saving mode 62 11 Maintenance 64 Servicing intervals 64 Cleaning 64 Avoiding staining 65 Checking and...

Страница 5: ...following functions Accessing the device status and program progress Accessing the user manuals and video tutorials for using the device Locating malfunctions and accessing solutions and assistance im...

Страница 6: ...operator s attention to the fact that they may be legally obliged e g in Germany ac cording to ElektroG to remove used batteries and accumulators non destructively before handing over the device provi...

Страница 7: ...eived for use outside the patient area The device should be located a minimum of 1 5 m radius away from the treatment area n The documentation media computer CF card reader etc must be placed in such...

Страница 8: ...pack the load and re sterilize it Transport and storage n Store and transport the device in a frost free environment n The device should always be carried by two people n Use suitable carrying straps...

Страница 9: ...ment cavities This permits the sterilization of large quantities of instruments or tex tiles directly one after each other thereby achieving excellent drying results Type of the feed water supply The...

Страница 10: ...tests Universal Program Quick Program B Quick Program S Gentle Program Prion Program Program type in accordance with EN 13060 Type B Type B Type S Type B Type B Dynamic pressure test of the sterilizat...

Страница 11: ...izing If the pressure and temperature correspond to the program dependent nominal values the sterilization phase begins The corresponding process parameters pressure and temperature are held at steril...

Страница 12: ...ies for small steam sterilizers Manufacturer s inspection report and declaration of conformity Warranty certificate Technical manual Record of installation and setup Tray lifter Power cable 4x cover c...

Страница 13: ...rgency 13 Front device foot adjustable 14 Manometer for pressure display on the double jacket steam generator behind cover 23 24 25 22 18 16 17 21 15 20 19 27 26 Fig 2 View from rear 15 Ethernet conne...

Страница 14: ...rvice hatch The service hatch is opened by pressing on the recess The service hatch is closed again by closing and pressing on the recess Fig 4 Opening and closing the service hatch Symbols on the dev...

Страница 15: ...ce may not be disposed as domestic waste The vendor is responsible for appropriate disposal of the device it must be delivered to the vendor to be disposed of MELAG devices are synonymous for long ter...

Страница 16: ...ivated during a program run State Meaning Lighted The energy saving mode can be activated Not lighted The energy saving mode can not be activated Colour touch display The operating panel consists of a...

Страница 17: ...fo status window Displays information regarding the software version and device data e g total number of batches maintenance counter log settings log memory and further technical values Service area O...

Страница 18: ...indicates different situations with various colours Colour of the LED Meaning Blue Standby program running drying has not yet begun Green Drying running program completed successfully Yellow Warning...

Страница 19: ...of feed water is sufficient for up to 25 steriliza tion runs Fill the storage container with feed water The water level of the storage container may not fall below the MIN mark during operation Check...

Страница 20: ...on the device type it requires a heat up time of approx 9 13 min after being switched on This time is required for the pre heating of the double jacket steam generator Opening and closing the door Th...

Страница 21: ...the door closed for min 3 s until the door lock engages Manual door emergency opening CAUTION Danger of scalding from hot steam Hot steam egress from the sterilization chamber is possible e g if it is...

Страница 22: ...y included in the scope of delivery in the opening The Allen key can be stored in the bracket behind the service hatch intended for this purpose 4 Turn the Allen key to open the door in a clockwise di...

Страница 23: ...gents Do not use any water repellent agents or oils impermeable to steam u When using ultrasonic devices care equipment for handpieces and washer disinfectors comply with the manufacturer s reprocessi...

Страница 24: ...packaging is important in achieving successful sterilization results You can use re usable rigid packaging systems or soft packaging such as transparent sterilization packaging paper bags sterilizati...

Страница 25: ...the seam seal tears during sterilization this could be caused by the choice of undersized packaging Re pack the instruments with larger packaging and perform sterilization again u Should the seam sea...

Страница 26: ...efore starting work the steam sterilizer must be cold and dry Batch related tests With Critical B instruments MELAcontrol Helix MELAcontrol Pro must be used as batch control with every sterilization c...

Страница 27: ...ter connection is present and the mains voltage Drying Activation of the intelligent drying subjects the drying phase to automatic managing and ends as soon as the load is dry Program name Packaging E...

Страница 28: ...ram see Additional drying page 55 Start time pre selection NOTICE Unsupervised operation of electrical devices including this steam sterilizer at the operator s risk MELAG accepts no liability what so...

Страница 29: ...rogram the door closes pressure tight and the steam sterilizer checks the quantity of feed water and its conductivity 1 To start a program press the START key 2 With activated user authentication Ente...

Страница 30: ...or or the system responding to an malfunction aborts the program run A system abort returns the steam sterilizer to a pressureless state This explains why a system abort takes longer than an abort by...

Страница 31: ...This could result in scalding n Use a tray lifter to remove the tray n Never touch the sterile material the sterilization chamber or the door with unprotected hands The components are hot Program abo...

Страница 32: ...d successful ly The steam sterilizer issues a malfunction message You then need to expect insufficient drying espe cially in the case of wrapped sterile material and a full load Sterile storage requir...

Страница 33: ...menu the log of the completed program will be outputted to the activated output medium after opening the door The approval process See also Approving the sterilization batch Video tutorial According...

Страница 34: ...team n Should the packaging be damaged or have burst after sterilization wrap the load again and re sterilize it If you remove the sterile material from the device directly after the end of the progra...

Страница 35: ...wing listed criteria u Comply with the maximum storage duration in accordance with the packaging type Comply with the manufacturer s information on the packaging u Do not store the sterile material in...

Страница 36: ...e is displayed Internal log memory full You now have the last chance to archive logs which have not been outputted confirm query with YES before the data in the steam sterilizer log memory up to the l...

Страница 37: ...ation pertaining to the requirements and setting of the computer as an output medium is available in the chapter Settings logging page 45 Reading out a text log on the computer All text logs can be op...

Страница 38: ...arked with a label to a specific batch the sterilization log file name must not be changed Type serial number Sterilization date Storage duration Daily batches no total batches User ID File name Stora...

Страница 39: ...ending Explanation Last log PRO The log of the last successful completed program is outputted Logs of the day PRO The log of the last successful program of the current day is outputted Logs of the we...

Страница 40: ...s that have been saved in the internal memory To facilitate the search you can filter the log sorting sequence by date program or outcome by selecting the top line 3 Select a log and press CONTINUE 4...

Страница 41: ...be saved Log directory A folder is created on all memory media CF card or computer after log output containing the encoded se rial number of the steam sterilizer concerned The folder name consists of...

Страница 42: ...5DA1 9A5EF5192C68174C868556542F7B8B05E97C6E46 16CDCFFA811E126FD67363FB74128A5F83AE6F37 F45A9E240C88615F1618D340060C1027205C83C2 Authentication of batch log 0 00 0 0 0 0 0 0 0 0 edk etm etd etp etv ett...

Страница 43: ...cumstances Once a week in routine operation During commissioning Following longer operating pauses Following a malfunction e g in the vacuum system Perform the Vacuum test with the steam sterilizer in...

Страница 44: ...chamber of the steam sterilizer and close the door 3 Working in menu Programs Tests select Bowie Dick test and press START Evaluation of the indicator following the colour change Depending on the manu...

Страница 45: ...ut is not to be performed immediately following the end of a program but saved in the internal memory e g to be collected and outputted together within a single week you can deactivate the immedi ate...

Страница 46: ...o Immediate output 2 Press CONTINUE and select the CF card and or the computer as the output medium 3 If necessary change the intervals and press CONTINUE 4 Working in this window check whether at lea...

Страница 47: ...d Log output in English If you want to print all text logs on the MELAprint log printer in English proceed as follows The text log is to be printed in English regardless of the language of the graphic...

Страница 48: ...fected via FTP or TCP Connection via FTP An FTP server or an FTP service is installed on the computer 1 Select Connection via FTP The lower pushbutton displays the current user data settings standard...

Страница 49: ...f the network technology Errors in the handling of IP addresses can result in malfunctions and data loss in your practice network n IP addresses may only be set by the practice network system administ...

Страница 50: ...2 Working in the logging assistant navigate to the window in which the IP addresses of the individual devices are listed 3 Select e g the steam sterilizer autoclave The settings window opens 4 Select...

Страница 51: ...inistration An ID and individual user PIN can be issued to every user with which to authenticate him herself so as to enable reliable traceability via the clearance process You can determine the neces...

Страница 52: ...e settings with SAVE then leave the menu 7 Exit the menu by pressing this symbol Deleting a user 1 Select the User administration option as described above and open the user list 2 Select the user ID...

Страница 53: ...uthentication upon program start Query user authentication upon program end Query user authentication upon program start and end You can skip the query user authentication Determining options for the...

Страница 54: ...start or after program end Press the key to skip the user authentication 6 Accept all the settings with SAVE then leave the menu Formatting the CF card NOTICE All data saved on the CF card is deleted...

Страница 55: ...s to the settings window 2 Press pushbutton YES or NO to choose whether additional drying should be performed during all subsequent program runs 3 Confirm with SAVE Activating deactivating additional...

Страница 56: ...ligent drying The display switches to the corresponding window 2 If you wish to deactivate intelligent drying select NO 3 Confirm with SAVE Date and time Correct batch documentation requires the corre...

Страница 57: ...pushbutton or to adjust the brightness and contrast on the display 3 Accept all the settings with SAVE and then leave the menu Volume 1 Select menu Settings Volume The display switches to the setting...

Страница 58: ...ings window 2 Press the CLASSIC button The design changes immediately 3 Press CONTINUE 4 Tap on a colour box to change the background colour e g blue The background colour changes immediately and the...

Страница 59: ...utomatically to the Settings menu MELAconnect The MELAconnect app is used to monitor the preparation processes of your MELAG steam sterilizers on the mobile end device e g smartphone tablet The follow...

Страница 60: ...e display to check whether MELAconnect has been connected to the correct device Key tone 1 Select menu Settings Key tone The display switches to the settings window 2 Press YES or NO to determine whet...

Страница 61: ...ttons 3 Confirm the settings with SAVE The display switches automatically to the Settings menu Explanation of the slide show options Display duration per picture indicates the time in seconds between...

Страница 62: ...et waiting time of 15 min the temperature of the double jacket steam generator falls to 103 C The program run time increases by approx 2 min upon the next start Waiting time 2 W2 After a pre set waiti...

Страница 63: ...steam sterilizer is in energy saving mode waiting time 2 1 Set the checkmark next to Activated and set the number of seconds after which the display is to be deactivated 2 Confirm the settings with SA...

Страница 64: ...om mends weekly cleaning of the surfaces e g with the MELAG chamber cleaning set Cleaning the sterilization chamber chamber sealing face mount and trays The following must be fulfilled or present The...

Страница 65: ...s and in the sterilization chamber as deposits and stains All steam conducting parts of the device consist of non rusting material This rules out the possibility of stain or rust development being cau...

Страница 66: ...is vital to ensure reliable operation and value retention of the device All function and safety relevant components and electrical units must be checked during maintenance and replaced where necessary...

Страница 67: ...night or the weekend Open the door and switch off the steam sterilizer Leave the door ajar to prevent premature wear and the sticking of the door seal Longer than two weeks Perform a vacuum test Afte...

Страница 68: ...e carried by two people n Use the correct carrying straps to carry the device Transport within the practice Comply with the following provisions during transport within a room or the practice u Decomm...

Страница 69: ...nd practice team n Re pack the load if necessary n Repeat the sterilization of the load All messages with current descriptions can be found in the Troubleshooting portal on the MELAG website https www...

Страница 70: ...the pressure tank is sufficiently full with water e Check whether the water inflow tap for the water treatment unit is open When using a MELAdem 47 check whether the tap on the pressure tank is open W...

Страница 71: ...or sagging and at a constant decline b Check whether the building siphon is blocked Please note If multiple devices are operated simultaneously we recommend the installation of an additional siphon c...

Страница 72: ...emergency opening page 21 clean the door seal and the seal face on the sterilization chamber see Cleaning page 64 83 The door does not reach a pressure tight state after the program start a The door s...

Страница 73: ...off the steam sterilizer at the power switch during a program run b Check whether the power plug is connected the power cable has suffered damage or a loose contact or loose plug connections is the c...

Страница 74: ...etector is blocked by soiling in the building supply d Unplug the leakage water detector control device wait approx 30 s and plug it back in again A switching noise on the water inflow tap black box o...

Страница 75: ...vacuum pump has suffered a blockage e g following long shutdown periods b A vacuum pump can be unblocked in the following fashion 1 Acknowledge the malfunction message 2 Switch off the steam steriliz...

Страница 76: ...see event 102 203 No log output options have been set Check the configuration in the menu Settings Logging 204 The internal log memory is full Output the log saved in the steam sterilizer on any outpu...

Страница 77: ...the CF card is write protected and remove the write protection Upon repeated incidence transfer the present log files from the CF card to a computer and format the CF card in the steam sterilizer and...

Страница 78: ...ompatible c The computer is not switched on d The network connection was not configured correctly e The documentation software on the computer was not started a Check whether the network cable has bee...

Страница 79: ...Start it again The steam sterilizer is now ready for operation Should this repeat please contact the service technician 438 The steam sterilizer must be validated Arrange for validation of the steam s...

Страница 80: ...mp can be unblocked in the following fashion 1 Acknowledge the malfunction message 2 Switch off the steam sterilizer and disconnect the power cable 3 Remove the round plastic cover on the right hand f...

Страница 81: ...rounding conditions Noise emission 72 dB A Heat emission with max load 0 95 kWh Ambient temperature 5 40 C ideal range 16 26 C Degree of protection following IEC 60529 IP20 Relative humidity max 80 at...

Страница 82: ...2 17G 41 x 14 x 9 cm ME01173 23M 42 x 16 x 6 cm ME01231 23G 42 x 16 x 12 cm ME01232 28M 32 x 16 x 6 cm ME01284 28G 32 x 16 x 12 cm ME01285 MELAstore System MELAstore Tray 50 18 x 11 8 x 3 cm ME01180 M...

Страница 83: ...15 Accessories and spare parts 83 Category Article Art no Spare parts Oil for door lock nut ME27515 Test gauge TR16 for door lock nut ME27522 Door seal ME45160...

Страница 84: ...rticles in space Corrosion The chemical alteration or destruction of metal ma terials by water and chemicals Delay in boiling This refers to the phenomenon that it is possible under certain circumstan...

Страница 85: ...er wrapped in textiles Network cable crossover A crossover cable connects two computers through network cards directly without the use of a hub switch This type of connection corresponds with the netw...

Страница 86: ...A closed minimum packaging which prevents the entrance of micro organisms e g through sealing bags sealed and re usable containers and folded sterilization towels etc Sterile material Also referred t...

Страница 87: ......

Страница 88: ...e 6 10 10829 Berlin Germany Email info melag com Web www melag com Original instructions Responsible for content MELAG Medizintechnik GmbH Co KG We reserve the right to technical alterations BA_Evo_40...

Отзывы: