2 Installation requirements
6
Space requirements
Fig. 1: View from the front, the right and above
Dimensions
Vacuklav 23 B+
Vacuklav 31 B+
Width
A
42.5 cm
Height
B
49 cm
Depth, total
C
74 cm
62 cm
Clearance between the device feet
C
1
45 cm
37 cm
Clearance from rear device foot up to the rear panel
C
2
7 cm
8 cm
Min. clearance to the side
D
1
5 cm
Min. clearance to the side of the door hinge
D
2
10 cm
Min. clearance to the rear
E
10 cm
Tank lid (width)
F
23 cm
Tank lid (depth)
F
1
19.5 cm
Tank lid (clearance to the rear)
F
2
23 cm
Free area with a fully-opened door
G
66 cm
54 cm
Min. clearance to the top (can be pulled out / with exhaust
shaft)
--
5/20 cm
The area above the steam sterilizer should be freely accessible in order to enable easy filling of the
storage tank and to ensure good ventilation.