8 Maintenance
17
8 Maintenance
Checks and cleaning
CAUTION
The mounts can have sharp edges.
n
Wear suitable protective gloves.
NOTICE
Inappropriately performed cleaning can lead to the scratching of and damage to surfaces
and the development of leaks in sealing faces. This also favours the development of
soiling deposits and corrosion in the sterilization chamber.
Comply with all information regarding cleaning of the parts affected.
Check the sterilization chamber, including the door seal and the mounts for the load once a week for
impurities, deposits or damage. If you find any impurities, remove the trays or standard tray cassettes and
the mount from the sterilization chamber. Clean the soiled components. When cleaning the sterilization
chamber, the mount for the load and the door seal, comply with the following:
u
Switch off the device before cleaning and disconnect the power plug from the socket.
u
Ensure that the sterilization chamber is not hot.
u
Use a soft, non-fuzzing cloth.
u
Use a chlorine and vinegar-free cleaning agent.
u
Soak the cloth in cleaning alcohol or spirit and attempt to wipe away impurities.
u
Only if the sterilization chamber or mount has persistent soiling, use a mild stainless steel cleaning
agent with a pH value between 5 and 8.
u
Use a neutral liquid cleaning agent to clean the door seal.
u
Do not use any hard objects such as a metal saucepan cleaner or a steel brush.
u
Check the door seal for damage on a daily basis. Replace the door seal if necessary.
u
Check the contact pressure of the door. The entire surface of the door must be in contact with the
frame.