MELAG MELAseal 101 Comfortable Скачать руководство пользователя страница 3

 

Operating manual MELAseal

®

101

 

Page 1 

1. Installation 

1.1. Overview 

of 

appliance _______________________________________________________ 

 

Cutter  

Mounting provisions for working 
platform   

Pressure rail  

Knob (temperature) 

Sealing rail  

Lever with cylindrical knob  

4 Signal 

lamp 

   

1.2. 

Power cable and pressure lever  ______________________________________________

 

The power cable should be plugged into a safety 
socket with earthing. The pressure lever can either 

be inserted in the bottom right or the left bottom side 
of the appliance.

 

  

1.3. Sealing 

temperature_________________________________________________________

 

The MELA

seal

®

101 has a very precise regulation 

system for the sealing temperature, which is 
adjusted by turning the temperature knob (6). With 
the transparent packaging material such as 
MELA

fol

®

 most commonly used in autoclaves by 

doctors and dentists (one side paper - one side foil 
as described in DIN 58953 Part 4) the temperature 
knob should be set to the first third of the area 
marked with "MELA

fol

®

". Turning the knob anti-

clockwise lowers the heating power, turning 
clockwise increases it. In continuous operation 
(more than 20 seals consecutively), the sealing 
temperature should be rechecked so that the foil 
does not melt and the quality of the seals remains 

constant. As long as the same packaging material is 
being used, it is then not necessary to adjust the 
sealing temperature again during the sealing 
operations. A safety delay of 5 seconds between 
sealing impulses protects against overheating during 
continuous operation.  
The use of transparent sterilization packaging with 
the MELA

seal

®

 foil sealer leads to seals of a uniform 

quality. Since this uniformity cannot be guaranteed 
with almost all types of tubular foil and these also 
lead to excessive wear of the pressing and sealing 
rails, the use of the MELA

seal

®

101 is not 

recommended with these foils. 

1.4. Wall 

mounting 

_____________________________________________________________

 

Holes are provided for wall mounting of the 
MELAseal. These are 22 cm apart. The appliance 

should only be mounted on load-bearing walls using 
S 6 plugs.  

 

            

6

1

4

5

2

3

7

 

MELAseal

 

Содержание MELAseal 101 Comfortable

Страница 1: ...lization equipment for medical practice During this period MELAG has succeeded in becoming a leading manufacturer of sterilization equipment More than 335 000 MELAG units sold throughout the world tes...

Страница 2: ...packaging rolls 2 3 Accessories 3 3 1 Work platform 3 3 2 Unit mounted roll holder Comfort 3 3 3 Wall mounted roll holder 3 3 4 Acrylic hood 3 3 5 Standard roll holder 3 4 Servicing 3 4 1 Exchanging t...

Страница 3: ...to the first third of the area marked with MELAfol Turning the knob anti clockwise lowers the heating power turning clockwise increases it In continuous operation more than 20 seals consecutively the...

Страница 4: ...tion After the signal lamps have gone out indicating the end of the sealing process then before the next sealing impulse the lever must be returned to its initial vertical position 2 2 Packaging cutte...

Страница 5: ...slit in the base of the cupboard Three different positions can be selected depending on the width of the roll used in order to ensure that the packaging foil is fed properly to the package sealing de...

Страница 6: ..._________ Loosen the four fastening screws and unscrew the round black knob then lift up the cover from the pressure rail Undo the fastening screw of the cutting blade Warning The blade is very sharp...

Страница 7: ...nd the seam in order to ensure a perfect seal without wrinkles This applies both for bags and transparent material from the roll 5 2 Storage life ______________________________________________________...

Страница 8: ...that micro organisms may penetrate the sterile goods packaging provided that dry storage conditions are ensured and the packaging material has the required germ proof qualities but that when the steri...