Chapter 2 – Commissioning
14
Connecting the log printer to the MELAseal Pro
Should you wish to use the log printer MELAprint 42 as output medium, connect it as follows:
1.
Connect the MELAprint 42 log printer serial connection cable to the RS232 interface on the rear
of the sealing device.
2.
Connect the power supply cable included in the scope of delivery of the log printer in accordance
with the appropriate operating manual.
Setting the log printer MELAseal Pro as the output medium
The sealing device recognises the log printer automatically. The sealing device does not need to be set as
the output medium.
The log printer outputs a log row for every sealing procedure performed.
Connecting a log printer to a MELAG device with an Ethernet
data connection
Applies to the devices:
MELAtherm 10
Cliniclave 45/45 M
Premium-Plus-Class / Evolution
Premium-Class
Vacu
quick
13-B, 14-B
Connecting the network adapter
PLEASE NOTE
A network adapter (art. no 40295) is required to connect a MELAG device with an Ethernet data
connection to the log printer.
Fig. 11: Connecting the network adapter
(1)
DC connection log printer
(2)
RS232 connection
(3)
Network adapter
(4)
DC connection network adapter
(5)
RJ12 connection
(6)
RJ45 connection
First connect the network adapter to the log printer:
1.
Insert the serial cable to the RS232 data connection of the log printer.
2.
Connect the other end of the serial cable to the RJ12 socket of the network adapter.
3.
Connect the network adapter to your MELAG device using the Ethernet data connection (RJ45
socket). The operating manual of the respective MELAG device contains information as to the
location of the Ethernet data connection.
3
2
1
6
5
4
Содержание MELAprint 42