2 Installation requirements
10
Operating the device with memory media
To prevent data loss, only use memory media to save the log data with the following characteristics:
n
Functional capability (without malware etc.)
n
Writeable
n
Formatted with a correct file system
Perform regular data backup. Restrict access to the device and systems with access authorization to the necessary circle
of persons.
Only use MELAG USB flash drives.
Operating the device in the local network (LAN)
NOTICE
Do not connect the device to a public network (e.g. the internet).
An Ethernet/IP-based network connection (LAN) is required to operate the device in a local network. In its delivery state,
the device is configured to obtain the IP address automatically from a DHCP server operated in a LAN.
NOTICE
Check the IP address carefully during the conversion for a manual configuration before connecting
the device to the LAN.
An incorrectly-entered IP address can cause IP conflicts in the network and thus disturb another device in your
network.
In the LAN with a firewall, only permit connections to and from the device which correspond to the intended use of the
device. All ports not used are blocked on the device side.
The device is able to make the following connections as standard:
Log
Source port
Destination
port
Direction
Aims
TCP
63000 - 64000 21
Outgoing
FTP control
TCP
any
63000 - 64000
Listening /
incoming
FTP (passive) data transfer (device set to
FTP logging)
UDP
68
67
Outgoing
Communication to DHCP server - requests
to the DHCP server
UDP
67
68
Listening /
incoming
Answers from DHCP server(s)
TCP
any
3333
Listening /
incoming
Data transfer log data (device set to TCP
logging)
UDP
62000
3000
Outgoing
Broadcast search label printer
UDP
3000
62000
Listening /
incoming
Search results label printer
TCP
≥ 1025
9100
Outgoing
Data transfer to the label printer