2 Safety instructions for building-in equipment
This building-in equipment has been designed and
tested in accordance with the requirements of Stan-
dard IEC 60950-1 "Safety of Information Technology
Equipment, including Electrical Business Equipment".
During installation of the building-in equipment in
an end application (i.e. rack) additional requirements
in accordance with Standard IEC60950-1 have to be
taken into account.
NOTE: First attach the case to protective earth -
before you connect the M4000 with the power line
(see chapter Grounding connection M4000).
General Safety instructions
•
The building-in equipment has been evaluated for use in office environment (pollution degree 2) and
may be only used in this environment. For use in rooms with a higher pollution degree more stringent
requirements are applicable.
•
The equipment/building-in equipment was evaluated for use in a maximum ambient temperature of 40
◦
C.
•
The building-in equipment may not be opened.
•
Protection against fire must be assured in the end application.
•
The ventilation opening may not be covered.
For AC Supply 100-240 V AC
•
The building-in equipment is a class 1 - equipment and must be connected to an earthed outlet
(TN Power System).
•
For safe operation the building-in equipment must be protected by max 16 A fuse in the power installation
system.
•
Disconnection of the equipment from mains is done by pulling the mains plug at the outlet. Don’t use the
connector at the module for disconnection from mains.
For DC Supply 100-200 V DC
•
The device can be disconnected outside the unit in accordance with the regulations as in IEC 60950-1
(e.g. through primary side line protection).
•
Assembling and disassembling of the power connector is only allowed if the device is disconnected from
power supply (e.g. trough primary side line protection).
•
All feed lines are sufficiently protected and dimensioned.
Fuse:
T2.5A
Connector Diameter:
1mm
2
- 2,5mm
2
/ 17AWG - 13AWG
2
Date: 21st August 2019
IMS-M4000