14
TORKEL 900-SERiES
ZP-CS01E
CS033663DE
4 MENU SYSTEM
4
M
enu system
4.1 Main menu
Screen buttons
Confirm
Exit without changes
Run
Pause the test to make any cor-
rections/changes in the settings*
)
Stop
Temperature compensation
Shown when compensation is
enabled
Quick guide
Open test file library
New test
Test configuration
information bar
Shows messages and is used for
acknowledge alarms
information field
Shows voltage and current
*
)
The pause time has limitations regulated by inter-
national standards. if the pause time is too long the
battery may have to be recharged before a new test is
done.
TEST menu
Test method
A test can be made in six test methods, "Constant i"
is default.
1]
Press the button and select test method.
▪
Constant i (current)
▪
Constant P (power)
▪
Constant R (resistance)
▪
Profile i (current profile)
▪
Profile P (power profile)
▪
V Logger (Current with monitoring during charge)
A]
For
Constant I, P or R
you press the numeric
field and make the settings for the parameter
using the On-screen keyboard or the control
knob.
B]
For
Profile I
or
Profile P
a new tab will be
shown on top of the screen together with a
window for the settings.
Here you configure your profile and press
to
start the test or press
TEST
to return to the test menu.
C]
For
V Logger
the voltage will be logged dur-
ing charge. You can set time values for warn-
ing and stop.
Warning and stop limit parameters
You can set TORKEL to issue a warning and/or to stop:
▪
When the voltage has reached a certain level.
▪
When a certain amount of capacity is discharged.
▪
After a specified time.
▪
When a cell voltage has reached a certain level (if BVM is
used, see section "9.2 BVM - Battery Voltage Monitor" on
page 37).
Содержание TORKEL 910
Страница 2: ......
Страница 13: ...CS033663DE ZP CS01E TORKEL 900 series 13 4 Menu system...
Страница 27: ...CS033663DE ZP CS01E TORKEL 900 series 27 6 External current measurement...
Страница 31: ...CS033663DE ZP CS01E TORKEL 900 series 31 7 Alarm function...
Страница 33: ...CS033663DE ZP CS01E TORKEL 900 series 33 8 How to obtain the desired current...
Страница 41: ...CS033663DE ZP CS01E TORKEL 900 series 41 10 Troubleshooting...
Страница 43: ...CS033663DE ZP CS01E TORKEL 900 series 43 11 Calibration...
Страница 45: ...CS033663DE ZP CS01E TORKEL 900 series 45 12 Specifications...
Страница 49: ...CS033663DE ZP CS01E TORKEL 900 series 49 12 Specifications...
Страница 51: ......